15
14
FUNCTIONAL DESCRIPTION HI 931000
FUNCTIONAL DESCRIPTION HI 931000
FUNCTIONAL DESCRIPTION HI 931000
FUNCTIONAL DESCRIPTION HI 931000
FUNCTIONAL DESCRIPTION HI 931000
ins rum n s
CFM
°C
CAL
OFF
ON
HI 931000
1
2
3
4
5
6
1) DIN socket for combination pH/°C electrode.
2) LCD display.
3)
ON/OFF
to switch the instrument on or off.
4)
°C
to display the measured temperature (press and hold).
5)
CAL
to enter calibration mode.
6)
CFM
to confirm calibration data.
SPECIFICATIONS HI 8424
SPECIFICATIONS HI 8424
SPECIFICATIONS HI 8424
SPECIFICATIONS HI 8424
SPECIFICATIONS HI 8424
HI 8424
Range
pH
0.00 to 14.00
ISE
0.0 to ±399.9 mV
ORP
±400 to ±1999 mV
°C
0.0 to 100.0
Resolution
pH
0.01
ISE
0.1mV
ORP
1mV
°C
0.1
Accuracy
pH
±0.01
(@20°C/68°F) ISE
±0.2mV
ORP
±1mV
°C
±0.4
Typical EMC
pH
±0.02
Deviation
ISE
±0.2
ORP
±1
°C
±0.4
pH Calibration
Automatic 2 point with
3 memorized standard buffers
(pH 7.01, 4.01 or 10.01)
Offset Calibration
±1.0 pH
Slope Calibration
from 70 to 108%
Temperature
Automatic from 0 to 100°C (32 to 212°F)
Compensation
or manual without temperature probe
Electrode
HI 1230B
combination, double-junction,
gel pH electrode BNC
and 1 m (3.3') cable (
included
)
Temperature
HI 7669AW
(
included
)
Probe
Input Impedance
10
12
ohm
Battery Type
1 x 9 volt
Life
100 hours of continuous use
Environment
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Dimensions
185 x 82 x 45 mm
(7.3 x 3.2 x 1.8")
Weight
520 g (1.1 lb.)