43

42

BATTERY REPLACEMENT

BATTERY REPLACEMENT

BATTERY REPLACEMENT

BATTERY REPLACEMENT

BATTERY REPLACEMENT

All meters are powered by a 9V battery that is located on the rear of
the instrument.
When the battery becomes weak:
-

HI 8014

, reaching a certain threshold, automatically switches the

LCD off.

-

HI 8314

,

 HI 8915

,

 HI 8424

 display a blink-

ing additional decimal point on the left side
of the LCD.

-

HI 9214

 and 

HI 931000

 display "Eb".

-

HI 9622

 displays "BAT" on the left side of

the LCD.

When the low battery indicator appears only a few hours of battery
life is remaining. A low battery may also result in loss of calibration
data or unreliable measurements. It is recommended that the battery
be replaced immediately.
Replacement must only take place in a non-hazardous area using an
alkaline 9V battery.
To access the battery, remove the battery cover by applying pressure
in the direction indicated below. Replace the old battery with a new
one while paying attention to its polarity.

Note

: the instrument should be recalibrated after changing the

battery.

HI 8114
HI 8414
HI 9214

9 VOLT BATTERY

SLIDE

HI  9214

HI 8014 - HI 8314
HI 8424 - HI 8915

HI 9622

HI  931000

HI 8014
HI 8314
HI 8424
HI 8915

BATTERY

BAT

pH

Note:

for field applications, it is always recommended to keep a
spare electrode handy. When anomalies are not resolved with
a simple maintenance, change the electrode (and recalibrate
the meter) to see if the problem is alleviated.

TAKING REDOX MEASUREMENTS

TAKING REDOX MEASUREMENTS

TAKING REDOX MEASUREMENTS

TAKING REDOX MEASUREMENTS

TAKING REDOX MEASUREMENTS

HI 8014

HI 8314

HI 8424

HI 8915

 and 

HI 9622

 have the

capability to take ORP measurements.
An optional ORP electrode must be used to perform these measure-
ments.
Oxidation-reduction potential (REDOX) measurements provide the
quantification of the oxidizing or reducing power of the sample tested.
To correctly perform a redox measurement, the surface of the ORP
electrode must be clean and smooth.
When not in use, the tip of the electrode should be kept moist and
safe from any mechanical stress which might cause damage to the
glass/platinum junction.

Содержание HI 8014

Страница 1: ...Instruction Manual HI 8014 HI 8314 HI 8424 HI 8915 HI 9214 HI 931000 HI 9622 Portable pH Meters h t t p w w w h a n n a i n s t c o m These Instruments are in Compliance with the CE Directives...

Страница 2: ...panel HI 8314 comes supplied with HI 1230B combination double junction gel pH electrode HI 7669AW temperature probe a 9 volt battery and a calibration screwdriver HI 8424 is a portable microprocessor...

Страница 3: ...HI 8915 comes supplied with HI 1230B combination double junction gel pH electrode HI 7669AW temperature probe a 9 volt battery and a calibration screwdriver HI 9214 is a compact microprocessor based...

Страница 4: ...4 9 6 pH ON OFF mV C pH HI 8314 membrane pHmeter STD 1nd Calibration at pH 6 86 or 7 01 SLOPE 2nd Calibration at pH 4 01 or 10 01 CALIBRATION HI 8014 Range pH 0 00 to 14 00 mV 0 to 1999 Resolution pH...

Страница 5: ...ive connection pH ON OFF HOLD mV C pH HI 8915 ATC pHmeter mV OUTPUT SLOPE 4 01 10 01 OFFSET 6 86 7 00 CALIBRATION 1 7 2 9 4 10 5 11 8 3 6 12 SPECIFICATIONS HI 8314 SPECIFICATIONS HI 8314 SPECIFICATION...

Страница 6: ...0 1 Accuracy pH 0 01 20 C 68 F mV 1 C 0 5 Typical EMC pH 0 02 Deviation mV 1 C 0 8 pH Calibration Manual 2 point through offset and slope trimmers Offset Calibration 1 pH Slope Calibration from 85 to...

Страница 7: ...irm calibration data HI 9214 Range pH 0 00 to 14 00 C 0 0 to 70 0 Resolution pH 0 01 C 0 1 Accuracy pH 0 02 20 C 68 F C 0 5 Typical EMC pH 0 08 Deviation C 1 pH Calibration Automatic 2 point with 3 me...

Страница 8: ...00 ISE 0 0 to 399 9 mV ORP 400 to 1999 mV C 0 0 to 100 0 Resolution pH 0 01 ISE 0 1mV ORP 1mV C 0 1 Accuracy pH 0 01 20 C 68 F ISE 0 2mV ORP 1mV C 0 4 Typical EMC pH 0 02 Deviation ISE 0 2 ORP 1 C 0...

Страница 9: ...l compensation when the temperature probe is disconnected 9 CFM to confirm calibration data HI 931000 Range pH 0 00 to 14 00 C 0 0 to 70 0 Resolution pH 0 01 C 0 1 Accuracy pH 0 01 20 C 68 F C 0 5 Typ...

Страница 10: ...9622 the display will show the previ ous calibration date month date for 3 seconds For HI 9214 Connect the pH electrode to the screw type connector on the meter Be sure that the electrode and the met...

Страница 11: ...ional temperature probe is needed To use the Automatic Temperature Compensation feature of HI 8314 HI 8424 HI 8915 and HI 9622 submerge the temperature probe into the sample as close to the electrode...

Страница 12: ...E C compensate for it HI 8424 and HI 9622 only In this case the temperature probe must be disconnected With HI 8314 and HI 8915 when the temperature probe is discon nected the pH reading will be compe...

Страница 13: ...HI 8014 HI 8314 and HI 8915 to NBS standards use pH 6 86 HI 7006 or HI 8006 instead of pH 7 01 and pH 9 18 HI 7009 or HI 8009 instead of pH 10 01 PROCEDURE FOR HI 8014 Remove the protective cap from...

Страница 14: ...st the SLOPE trimmer on the lower right of the front panel until the LCD shows the pH value at the noted temperature The pH calibration is now complete HI 7007 4 cm 1 C C pH HI 7004 OFFSET pH SLOPE pH...

Страница 15: ...on the previous page HI 7004 pH pH CFM RANGE C HI 7007 C PROCEDURE FOR HI 8424 Switch the meter on after connecting the pH electrode and the temperature probe Remove the protective cap from the elec t...

Страница 16: ...first buffer solution ins e rum n s ON OF F CA L CF M HI 921 4 stic k pHm eter HI 7004 ON OFF CAL CFM HI 7007 4 cm 1 CFM PROCEDURE FOR HI 9214 Turn the meter on Note the stability indicator blinks wh...

Страница 17: ...offset one point calibration For best accuracy however it is recommended that a two point calibra tion is performed The LCD will now blink 4 01 pH 10 01 can be selected by pressing the CAL key Rinse w...

Страница 18: ...03 or 10 06 If the buffer temperature is 50 C the display should show pH 4 06 or 6 98 or 9 82 pH CAL The meter will request pH 4 01 or 10 01 buffer showing Er 4 This indicates that the instrument is...

Страница 19: ...libration An out of order pH electrode The electrode is not working or the refer ence junction is clogged Erroneous buffer so lution used for slope calibration An out of order pH electrode A clogged o...

Страница 20: ...ffer so lution used for slope calibration An out of order pH electrode Out of range pH scale No temperature probe attached Out of range mV scale Erroneous pH 7 buffer solution used for off set calibra...

Страница 21: ...ish the metal tip with a lightly abrasive paper paying attention not to scratch the surface and wash thoroughly with water If the bulb and or junction are dry soak the electrode in HI 70300 Storage So...

Страница 22: ...ention to its polarity Note the instrument should be recalibrated after changing the battery HI 8114 HI 8414 HI 9214 9 VOLT BATTERY SLIDE HI 9214 HI 8014 HI 8314 HI 8424 HI 8915 HI 9622 HI 931000 HI 8...

Страница 23: ...ning Solution 460 mL HI 7073M Protein Cleaning Solution 230 mL HI 7073L Protein Cleaning Solution 460 mL DISPLAY CODE GUIDE DISPLAY CODE GUIDE DISPLAY CODE GUIDE DISPLAY CODE GUIDE DISPLAY CODE GUIDE...

Страница 24: ...ution 230 mL HI 7077L Oil Fat Cleaning Solution 460 mL ELECTRODE CLEANING SOLUTIONS IN FDA APPROVED BOTTLE HI 8061M General Cleaning Solution 230 mL HI 8061L General Cleaning Solution 460 mL HI 8073M...

Страница 25: ...4 7 12 mm 0 5 FC 210 120 mm 4 7 12 mm 0 5 9 5mm DIA 0 37 FC 220 110 mm 4 3 12 mm 0 5 FC 911 HI 1330B HI 1310S Glass body semimicro single junction refillable combination pH electrode Use laboratory H...

Страница 26: ...931001 pH and ORP electrode simulator with LCD and 1 m 3 3 coaxial cable ending in female BNC connectors HI 710007 Shockproof blue rubberboot for HI 931000 formerly HI 935BTB HI 710008 Shockproof oran...

Страница 27: ...read HI 2031B FC 200B HI2020S FC 200S 5 Cheese FC 200B FC200S 6 Dairyproducts FC 911B FC 100B 7 Dirtywater HI 1910B HI 1912B 8 Emulsions HI 1053B HI1050S 9 Environment HI 1230B HI1210S 10 Flasks HI 13...

Страница 28: ...netic fields it is recommended to keep parallel the cables of the two probes and as close as possible to the instrument In particular cases HI 8915 could change its operating mode In such cases press...

Страница 29: ...ITERATURE HANNA LITERATURE LAB RECORDING WATER ANALYSIS HANDBOOK ENVIROCARE GENERAL CATALOG These and many others catalogs handbooks and leaflets are avail able from Hanna To receive your free copy co...

Отзывы: