background image

5

4

FUNCTIONAL DESCRIPTION HI 8014

FUNCTIONAL DESCRIPTION HI 8014

FUNCTIONAL DESCRIPTION HI 8014

FUNCTIONAL DESCRIPTION HI 8014

FUNCTIONAL DESCRIPTION HI 8014

1) BNC socket for combination pH electrodes.
2) LCD display.
3)

Rotary Knob 

with the following selectable positions:

OFF

To switch the meter off (in any other position the
instrument is on).

°C

To display the manual temperature setting (through
°C trimmer).

pH

To display the pH value.

mV

To display the mV (ORP) readings when using an
ORP electrode or the mV equivalent to the pH values
when using a pH electrode.

4)

Trimmers:
°C

To set temperature manually.

mV/pH

For SLOPE calibration of pH.

AP

For OFFSET calibration of pH.

ins rum n s

pH

pH meter

HI 8014

°C

mV/pH

AP

pH

°C

OFF

mV

2

1

3

4

pH 4.01, pH 7.01 and pH 10.01. It can also be used for ISE
and ORP measurements with automatic range switching when
readings exceed ±399.9 mV. 

HI 8424

 comes supplied with:

HI 1230B combination, double-junction, gel pH electrode,
HI 7669AW temperature probe and a 9 volt battery.

HI 8915

 is a portable 

pH/mV/°C meter

 that combines accurate

pH measurements with a recorder output capability and auto-
matic temperature compensation. Two recorder output terminals
make it possible to attach a plotter or recorder directly to the
meter for a hard copy of the measurements. Pressing the HOLD
key will freeze the reading on the display. Calibration is per-
formed using the two trimmers on the front panel. 

HI 8915

comes supplied with: HI  1230B combination, double-junction,
gel pH electrode, HI  7669AW temperature probe, a 9 volt
battery and a calibration screwdriver.

HI 9214

 is a compact microprocessor-based 

pH/°C meter

. The

electrode has a built-in temperature sensor for quick and accurate
temperature measurements and temperature compensation. The
meter has three memorized buffer values (pH 4.01, pH 7.01, pH
10.01) and automatic buffer recognition to avoid errors during
calibration. There are no trimmers making calibration an easy
process, even for nontechnical personnel. 

HI 9214

 comes supplied

with: HI 1217S screw-type combination gel pH/°C electrode, a 9
volt battery and HI 710003 soft carrying case.

HI 931000

 is a compact, microprocessor-based 

pH/°C

 

meter

 with

4-in-1 pH electrode (pH/°C electrode with built-in temperature
sensor and signal amplifier). Special shockproof rubberboots are
also available to prolong the life of the meter and to prevent
damage (see page 51). 

HI 931000

 comes supplied with: HI 1217D

combination gel pH/°C electrode, with a built-in temperature
sensor and signal amplifier, a 9 volt battery and HI 710004 soft
carrying case.

HI 9622

 is a portable microprocessor-based 

pH/mV/°C  meter

which meets the 

GLP

 (Good Laboratory Practice) requirements

and memorizes the last calibration date. The meter has three
memorized buffer values: pH 4.01, pH 7.01 and pH 10.01. The
meter can also be used for ISE and ORP measurements with
automatic range switching when readings exceed ±399.9 mV.

HI 9622

 comes supplied with: HI  1230B combination, double-

junction, gel pH electrode, HI  7669/2W temperature probe and
a 9 volt battery.

Содержание HI 8014

Страница 1: ...Instruction Manual HI 8014 HI 8314 HI 8424 HI 8915 HI 9214 HI 931000 HI 9622 Portable pH Meters h t t p w w w h a n n a i n s t c o m These Instruments are in Compliance with the CE Directives...

Страница 2: ...panel HI 8314 comes supplied with HI 1230B combination double junction gel pH electrode HI 7669AW temperature probe a 9 volt battery and a calibration screwdriver HI 8424 is a portable microprocessor...

Страница 3: ...HI 8915 comes supplied with HI 1230B combination double junction gel pH electrode HI 7669AW temperature probe a 9 volt battery and a calibration screwdriver HI 9214 is a compact microprocessor based...

Страница 4: ...4 9 6 pH ON OFF mV C pH HI 8314 membrane pHmeter STD 1nd Calibration at pH 6 86 or 7 01 SLOPE 2nd Calibration at pH 4 01 or 10 01 CALIBRATION HI 8014 Range pH 0 00 to 14 00 mV 0 to 1999 Resolution pH...

Страница 5: ...ive connection pH ON OFF HOLD mV C pH HI 8915 ATC pHmeter mV OUTPUT SLOPE 4 01 10 01 OFFSET 6 86 7 00 CALIBRATION 1 7 2 9 4 10 5 11 8 3 6 12 SPECIFICATIONS HI 8314 SPECIFICATIONS HI 8314 SPECIFICATION...

Страница 6: ...0 1 Accuracy pH 0 01 20 C 68 F mV 1 C 0 5 Typical EMC pH 0 02 Deviation mV 1 C 0 8 pH Calibration Manual 2 point through offset and slope trimmers Offset Calibration 1 pH Slope Calibration from 85 to...

Страница 7: ...irm calibration data HI 9214 Range pH 0 00 to 14 00 C 0 0 to 70 0 Resolution pH 0 01 C 0 1 Accuracy pH 0 02 20 C 68 F C 0 5 Typical EMC pH 0 08 Deviation C 1 pH Calibration Automatic 2 point with 3 me...

Страница 8: ...00 ISE 0 0 to 399 9 mV ORP 400 to 1999 mV C 0 0 to 100 0 Resolution pH 0 01 ISE 0 1mV ORP 1mV C 0 1 Accuracy pH 0 01 20 C 68 F ISE 0 2mV ORP 1mV C 0 4 Typical EMC pH 0 02 Deviation ISE 0 2 ORP 1 C 0...

Страница 9: ...l compensation when the temperature probe is disconnected 9 CFM to confirm calibration data HI 931000 Range pH 0 00 to 14 00 C 0 0 to 70 0 Resolution pH 0 01 C 0 1 Accuracy pH 0 01 20 C 68 F C 0 5 Typ...

Страница 10: ...9622 the display will show the previ ous calibration date month date for 3 seconds For HI 9214 Connect the pH electrode to the screw type connector on the meter Be sure that the electrode and the met...

Страница 11: ...ional temperature probe is needed To use the Automatic Temperature Compensation feature of HI 8314 HI 8424 HI 8915 and HI 9622 submerge the temperature probe into the sample as close to the electrode...

Страница 12: ...E C compensate for it HI 8424 and HI 9622 only In this case the temperature probe must be disconnected With HI 8314 and HI 8915 when the temperature probe is discon nected the pH reading will be compe...

Страница 13: ...HI 8014 HI 8314 and HI 8915 to NBS standards use pH 6 86 HI 7006 or HI 8006 instead of pH 7 01 and pH 9 18 HI 7009 or HI 8009 instead of pH 10 01 PROCEDURE FOR HI 8014 Remove the protective cap from...

Страница 14: ...st the SLOPE trimmer on the lower right of the front panel until the LCD shows the pH value at the noted temperature The pH calibration is now complete HI 7007 4 cm 1 C C pH HI 7004 OFFSET pH SLOPE pH...

Страница 15: ...on the previous page HI 7004 pH pH CFM RANGE C HI 7007 C PROCEDURE FOR HI 8424 Switch the meter on after connecting the pH electrode and the temperature probe Remove the protective cap from the elec t...

Страница 16: ...first buffer solution ins e rum n s ON OF F CA L CF M HI 921 4 stic k pHm eter HI 7004 ON OFF CAL CFM HI 7007 4 cm 1 CFM PROCEDURE FOR HI 9214 Turn the meter on Note the stability indicator blinks wh...

Страница 17: ...offset one point calibration For best accuracy however it is recommended that a two point calibra tion is performed The LCD will now blink 4 01 pH 10 01 can be selected by pressing the CAL key Rinse w...

Страница 18: ...03 or 10 06 If the buffer temperature is 50 C the display should show pH 4 06 or 6 98 or 9 82 pH CAL The meter will request pH 4 01 or 10 01 buffer showing Er 4 This indicates that the instrument is...

Страница 19: ...libration An out of order pH electrode The electrode is not working or the refer ence junction is clogged Erroneous buffer so lution used for slope calibration An out of order pH electrode A clogged o...

Страница 20: ...ffer so lution used for slope calibration An out of order pH electrode Out of range pH scale No temperature probe attached Out of range mV scale Erroneous pH 7 buffer solution used for off set calibra...

Страница 21: ...ish the metal tip with a lightly abrasive paper paying attention not to scratch the surface and wash thoroughly with water If the bulb and or junction are dry soak the electrode in HI 70300 Storage So...

Страница 22: ...ention to its polarity Note the instrument should be recalibrated after changing the battery HI 8114 HI 8414 HI 9214 9 VOLT BATTERY SLIDE HI 9214 HI 8014 HI 8314 HI 8424 HI 8915 HI 9622 HI 931000 HI 8...

Страница 23: ...ning Solution 460 mL HI 7073M Protein Cleaning Solution 230 mL HI 7073L Protein Cleaning Solution 460 mL DISPLAY CODE GUIDE DISPLAY CODE GUIDE DISPLAY CODE GUIDE DISPLAY CODE GUIDE DISPLAY CODE GUIDE...

Страница 24: ...ution 230 mL HI 7077L Oil Fat Cleaning Solution 460 mL ELECTRODE CLEANING SOLUTIONS IN FDA APPROVED BOTTLE HI 8061M General Cleaning Solution 230 mL HI 8061L General Cleaning Solution 460 mL HI 8073M...

Страница 25: ...4 7 12 mm 0 5 FC 210 120 mm 4 7 12 mm 0 5 9 5mm DIA 0 37 FC 220 110 mm 4 3 12 mm 0 5 FC 911 HI 1330B HI 1310S Glass body semimicro single junction refillable combination pH electrode Use laboratory H...

Страница 26: ...931001 pH and ORP electrode simulator with LCD and 1 m 3 3 coaxial cable ending in female BNC connectors HI 710007 Shockproof blue rubberboot for HI 931000 formerly HI 935BTB HI 710008 Shockproof oran...

Страница 27: ...read HI 2031B FC 200B HI2020S FC 200S 5 Cheese FC 200B FC200S 6 Dairyproducts FC 911B FC 100B 7 Dirtywater HI 1910B HI 1912B 8 Emulsions HI 1053B HI1050S 9 Environment HI 1230B HI1210S 10 Flasks HI 13...

Страница 28: ...netic fields it is recommended to keep parallel the cables of the two probes and as close as possible to the instrument In particular cases HI 8915 could change its operating mode In such cases press...

Страница 29: ...ITERATURE HANNA LITERATURE LAB RECORDING WATER ANALYSIS HANDBOOK ENVIROCARE GENERAL CATALOG These and many others catalogs handbooks and leaflets are avail able from Hanna To receive your free copy co...

Отзывы: