Care & Maintenance

10.  CARE & MAINTENANCE

•  Do not immerse the tester over the maximum immersion level.
•  Fresh buffer should be used for each calibration. Once the sachets are opened the buffer value can change 

over time.

•  For improved accuracy a two-point calibration is recommended.
•  If the electrode is slow or sluggish, soak it in cleaning solution for 20 minutes. Rinse with water and 

hydrate the electrode in storage solution for a minimum of 30 minutes before calibrating.

•  If measurements are taken successively, rinse the electrode thoroughly in distilled or deionized water to 

eliminate cross-contamination between measurements.

•  When not in use, remove the blade, add a few drops of storage solution to the protective cap to keep the glass 

tip and the junction hydrated. If storage solution is not available, pH 4.01 or pH 7.01 buffer can be used. 

Cleaning

•  Remove the blade and o-ring, and clean with soapy water.

•  Rinse with deionized water. Keep the o-ring with the blade after cleaning.

•  Remove electrode’s outer junction sleeve and clean with purified water. Once cleaned a small amount 

of supplied gel electrolyte should be added to refresh the junction and improve the pH measurement.

Note

: Never immerse the tester over the maximum immersion level.

Refilling the Electrode

•  To remove the electrode sleeve, carefully rotate and slide it off (A).

Set aside.

Note

: Handle the probe with care, the pH electrode stem is made of glass.

•  Rinse off any traces of electrolyte gel from sleeve. Soak the electrode tip in 

HI700601

 General purpose cleaning solution (or 

HI700661

HI700663

HI700664

) for 20 minutes. Rinse with distilled or deionized water.

•  Refill the reference well with 

HI9071

 Gelled bridge electrolyte (B).

•  Replace the sleeve. Ensure the o-ring is fixed inside the electrode. Rinse off 

excess gel with distilled or deionized water.

•  Shake the probe down gently to eliminate any trapped air bubbles.
•  Soak the electrode in 

HI70300

 Electrode storage solution for a minimum of 30 minutes before calibrating.

Storage

•  To ensure a quick response, the glass tip and junction should be kept hydrated.

•  When not in use, add a few drops of storage solution to the storage cap. If storage solution is not available, 

pH 4.01 or pH 7.01 buffer can be used.

Note

: Do not store the electrode in distilled or deionized water.

Stainless steel blade

•  When not in use, keep the 

FC097

 meat blade dry and clean (no grease or food particles) in the designated 

box slot, with protective cap on, to avoid contamination. Keep the o-ring with the blade.

A

B

Содержание HALO2 HI9810452

Страница 1: ...INSTRUCTION MANUAL Hanna Instruments Inc 584 Park East Drive Woonsocket RI 02895 USA www hannainst com HI9810452 Wireless pH Tester for Meat with built in specialized electrode Foodcare...

Страница 2: ...8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 8 Stand Alone Tester Up to Three Point Calibration 8 With Hanna Lab App Up to Four Point Calibration 8...

Страница 3: ...05 pH mV Range pH mV conversion Resolution 0 1 or 1 mV Temperature Range 0 0 to 60 0 C 32 0 to 140 0 F Resolution 0 1 C 0 1 F Accuracy 0 5 C 0 9 F Calibration Up to three points or four points Automa...

Страница 4: ...e g solvents sodium hypochlorite ultraviolet light and fungal growth the PVDF body has high abrasion resistance and mechanical strength The conical tip allows for easy penetration into solids and sem...

Страница 5: ...nitialization screen displays all LCD segments followed by the battery percentage The tester enters measurement mode Before the tester turns off OFF PWR is displayed briefly Press and hold the ON OFF...

Страница 6: ...ff 8 AOFF 60 AOFF AOFF Calibration Points Option 2P or 3P Press ON OFF button to select between up to two or up to three point calibration 2P CAL 3P CAL Buffer Set Option 7 01 pH Hanna or 6 86 pH NIST...

Страница 7: ...t of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the application Tag a Measurement Once connected to the applicatio...

Страница 8: ...return to measurement mode press the ON OFF button CAL SAVE is displayed briefly With 2P option selected 6 Place the tip of the electrode in the second buffer rinse beaker then in the second calibrat...

Страница 9: ...and improve the pH measurement Note Never immerse the tester over the maximum immersion level Refilling the Electrode To remove the electrode sleeve carefully rotate and slide it off A Set aside Note...

Страница 10: ...r solution 20 mL sachet 25 pcs HI70007P pH 7 01 buffer solution 20 mL sachet 25 pcs HI70009P pH 9 18 buffer solution 20 mL sachet 25 pcs HI70010P pH 10 01 buffer solution 20 mL sachet 25 pcs HI77400P...

Страница 11: ...and maintained according to instructions This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of prescribed maintenance is not covered If...

Страница 12: ...C 12208A 04 This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must...

Отзывы: