background image

Calibration 8

Procedure

For most applications it is recommended to start with pH 7.01 buffer (or pH 6.86).
To restore factory defaults, press and hold the ON/OFF button. “

CAL

 

CLR

” is displayed.

Note

: It is recommended to calibrate the electrode with buffers at the temperature it will be used at.

Stand-Alone Tester (Up to Three-Point Calibration)

1.  Rinse the electrode tip with purified water and blot dry. Then rinse with the buffer being used for calibration.
2.  Press and hold ON/OFF button until “

CAL

 

MODE

” is displayed.

3.  When “

7.01 

USE

” or “

6.86 

USE

” is displayed with “CAL” tag blinking, place the tip of the electrode 

in the correct buffer.

CAL

MODE

   

7.01

USE

4.  When the buffer has been recognized, “REC“ is displayed. 

WAIT

” is displayed with the stability indicator (

#

) blinking until the reading is stable. Wait until the 

measurement is stored and the stability indicator disappears.

 

7.01

REC

   

#

7..01

WAIT

5.  To save a one-point calibration and return to measurement mode, press the ON/OFF button.

CAL

 

SAVE

” is displayed briefly.

With 2P option selected

6.  Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait 

until the measurement is stored and the stability indicator disappears. 

7.  After the second point has been stored, “

CAL

 

SAVE

” is displayed briefly and the tester will return to 

measurement mode automatically.

With 3P option selected 

6.  Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait 

until the measurement is stored and the stability indicator disappears. 

7.  Place the tip of the electrode in the third buffer rinse beaker, then in the third calibration buffer. Wait 

until the measurement is stored and the stability indicator disappears. 

8.  After the third point has been stored, “

CAL

 

SAVE

” is displayed briefly and the tester will return to 

measurement mode automatically.

With Hanna Lab App (Up to Four-Point Calibration)

Connect the tester to the Hanna Lab App and follow the calibration procedure.
See the App Help section for the calibration procedure.

Содержание HALO2 HI9810452

Страница 1: ...INSTRUCTION MANUAL Hanna Instruments Inc 584 Park East Drive Woonsocket RI 02895 USA www hannainst com HI9810452 Wireless pH Tester for Meat with built in specialized electrode Foodcare...

Страница 2: ...8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 8 Stand Alone Tester Up to Three Point Calibration 8 With Hanna Lab App Up to Four Point Calibration 8...

Страница 3: ...05 pH mV Range pH mV conversion Resolution 0 1 or 1 mV Temperature Range 0 0 to 60 0 C 32 0 to 140 0 F Resolution 0 1 C 0 1 F Accuracy 0 5 C 0 9 F Calibration Up to three points or four points Automa...

Страница 4: ...e g solvents sodium hypochlorite ultraviolet light and fungal growth the PVDF body has high abrasion resistance and mechanical strength The conical tip allows for easy penetration into solids and sem...

Страница 5: ...nitialization screen displays all LCD segments followed by the battery percentage The tester enters measurement mode Before the tester turns off OFF PWR is displayed briefly Press and hold the ON OFF...

Страница 6: ...ff 8 AOFF 60 AOFF AOFF Calibration Points Option 2P or 3P Press ON OFF button to select between up to two or up to three point calibration 2P CAL 3P CAL Buffer Set Option 7 01 pH Hanna or 6 86 pH NIST...

Страница 7: ...t of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the application Tag a Measurement Once connected to the applicatio...

Страница 8: ...return to measurement mode press the ON OFF button CAL SAVE is displayed briefly With 2P option selected 6 Place the tip of the electrode in the second buffer rinse beaker then in the second calibrat...

Страница 9: ...and improve the pH measurement Note Never immerse the tester over the maximum immersion level Refilling the Electrode To remove the electrode sleeve carefully rotate and slide it off A Set aside Note...

Страница 10: ...r solution 20 mL sachet 25 pcs HI70007P pH 7 01 buffer solution 20 mL sachet 25 pcs HI70009P pH 9 18 buffer solution 20 mL sachet 25 pcs HI70010P pH 10 01 buffer solution 20 mL sachet 25 pcs HI77400P...

Страница 11: ...and maintained according to instructions This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of prescribed maintenance is not covered If...

Страница 12: ...C 12208A 04 This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must...

Отзывы: