Technical Support : +82-2-3452-8966 / [email protected] 5 /29
1. Safety Precautions
Before using the device, please read this operation manual to ensure correct usage. Incorrect
handling of the device may possibly result in personal injury or physical damage on the
device or its accessories.
1. ALWAYS locate the device on a flat, rigid and stable table capable of withstanding the
weight of the device and its spinning operation.
2. ALWAYS make a safety zone of 30 cm around the Device to indicate that neither
hazardous materials nor persons should be permitted within the area during operation.
▶ ALWAYS position the device with enough space on each side of device to ensure
proper air circulation.
3. ALWAYS install the device within a temperature and humidity controlled environment
(permissible ambient temperature: 5 ~ 35 °C, relative humidity: 30 ~ 85 %, atmospheric
pressure: 500 ~ 1,060 hPa).
4. Before connecting the power, check the rated voltage.
5. Should not use unapproved rotors and accessories.
▶ Only use rotors from Hanil Scientific Inc. with appropriate centrifugal tubes to embrace
sample containers tightly enough inside rotors.
6. Before operating the device, check if the rotor is securely fastened.
▶ Should operate the device with a rotor properly installed and secured to the motor
shaft.
7. Do not stop the rotor by hands during the device is running.
8. Emergency lid release should be performed only when rotor is completely stopped.
9. ALWAYS load the tubes symmetrically with evenly weighted samples to avoid rotor
imbalance. If necessary, use the water blank to counterbalance the unpaired sample.
10. The operation speed should not exceed the highest value of the individual guaranteed
g-forces of each centrifuge, rotor, sleeves and sample container, especially the guaranteed
g-force of sample container should not be neglected.
11. ALWAYS disconnect the power supply prior to maintenance and service to avoid electrical
shock.
1.3 Safety Precautions
Содержание HyperVAC HVC-2124