Technical Support : +82-2-3452-8966 / [email protected] 19/29
[Detailed Sequences for Operation]
1. Fill the chamber of Drip Catcher kit with ice.
2. Turn on the centrifuge and the vacuum pump.
3. Set the parameters according to your preferences. (RPM / TIME / TEMP)
4. Press the button then or button to switch the status of the vacuum pump.
Press the button upon setting each function. Normally, you don’t need to adjust the
values for these functions.
5. Press the button to start operation with your set values.
☞ If you set your speed above 1,000 rpm, the vacuum pump starts to operate at 1,000 rpm.
If you set it below 1,000 rpm, the vacuum pump starts to operate at your set RPM.
☞ Operation starts before the chamber reaches to the set temperature. If you want to start
the operation at your set temperature, you can use the PRE-HEAT function. Press the
button before pressing the button. Then the operation starts after reaching your set
temperature. LED blinks while setting and is kept on upon the completion of the setting.
Press the button to start operation.
6. If you want to stop the operation, press the button.
- Vacuum pump stops.
- Vacuum Release function is activated and the releasing is completed. Function LED blinks
during the release and is kept on upon the completion of the release.
- With the vacuum released, alarm beeps and Lid LED blinks for 5 seconds when the rotor
completely stops.
7. To open the lid, press the
button after the vacuum release. Then lid lock is released
for 5 seconds with the Lid LED on, and you may open it within the time.