www.ihanil.com
20/29
4. Operation
4.9 Manaul Operation
If you need to adjust the function of vacuum pump, to release it or execute emergency open,
you can use manual operation.
1. Turn on the main power.
2. Press the
button.
3. Check the display. (Default: Vacuum pump: OFF / Release: OFF)
4. Press the button and then press the or buttons to set the status of
the vacuum pump. You can switch on or off vacuum pump and release the vacuum on your
need. You need to press the button after setting each function.
5. If you want to execute the emergency lid open, the release has to be set as ‘ON’.
6. When the vacuum release is completed, alarm beeps and Lid LED blinks for 5 seconds.
7. Press the
button and open the lid within 5 seconds while the Lid LED blinks.
4.8 Operation Sequence
[Operating Status for Each Part (VacPump / Release / Heater)]
Power ON
(Start status: default setting)
VacPump
Release
Heater
OFF
OFF
OFF
Setting Temperature and Running
Device (~20 min.)
VacPump
Release
Heater
OFF
OFF
OFF
Until reaching the set RPM (Refer to
VacPump
Release
Heater
OFF
OFF
ON
After reaching the set RPM or 1,000 rpm
VacPump
Release
Heater
ON
OFF
ON
Press the STOP button.
VacPump
Release
Heater
OFF
ON
OFF
After completing the Stop –Return to the
Start Status (default setting)
VacPump
Release
Heater
OFF
OFF
OFF