background image

2. Installation

2-1. Safety

Centrifuge containing the high speed rotating part is dangerous. Follow

the instruction described in this manual to avoid breakage, trouble, injury.

1) Centrifuge must be placed the horizontally flat level, In the case of tilted

shaft of rotating big vibration occurs.

2) Check the voltage before connecting power, the improper volt may cause

the damage.

3) Use only the rotor, parts and accessories supplied by the maker, which is

not responsible for any damage caused by improper parts not

recommended by the maker.

4) Ascertain the material property completely and pre-arrange safety measure

in the case of handling pathogenic, toxic, radioactive material.

5) As Centrifuge is contaminated by pathogenic, toxic, radioactive material,

contaminant is to be cleared and ventilated and the isolation is needed.

6) Volatile and explosive material should not be treated by Centrifuge.

7) Don’t operate over the approved maximum rpm of rotor.
8) Don’t diminish and stop the rotating rotor with hand.

9) Rotor must be operated after symmetrical balance work of weight

10) Rotor must be fixed to the rotating shaft exactly and lid also fixed

completely. If not, the lid may open during operation.

11) Repairing and maintenance not mentioned on this manual must be

handled by the technician with the proper curriculum (in this case the

contaminant is to be cleared)

12) In order to keep the best condition and operation of long term the

chamber of rotor must be kept dry.

13) The chemical corrosion of rotor contacting strong acid, alkaline agent of

Cesium, silver, chlorine base should be avoided carefully.

Содержание FLETA 40P

Страница 1: ...Operation Manual FLETA 40P Multipurpose Centrifuge...

Страница 2: ...2 1 Safety 2 2 Location of Installation 2 3 Check list 2 4 Power Supply 3 Operation 3 1 Feature 3 2 Structure 3 3 Rotor 3 4 Function 3 5 Operating 4 Washing Maintenance 4 1 Washing 4 2 Maintenance 5...

Страница 3: ...s for safe operation therefore You must read in advance to the actual operation 1 1 Components Item Q nty Main body 1 Power cable 1 User Manual 1 Hexagonal T wrench 5mm 1 Packing check list 1 Check th...

Страница 4: ...minated by pathogenic toxic radioactive material contaminant is to be cleared and ventilated and the isolation is needed 6 Volatile and explosive material should not be treated by Centrifuge 7 Don t o...

Страница 5: ...ot be covered by the another instrument or fabric avoid the place of heavy dust Constant temp humidity The sensitive electronic controller is affected by ambient temp and humidity Avoid direct sunray...

Страница 6: ...d power supply to protect main body Door lock device Door sensor lets rotor stop in the case of open door and Centrifuge doesn t operate at the status of door open Auto door lock device doesn t allow...

Страница 7: ...ide of laboratory a remote area from Centrifuge to cut off power supply The safety distance 30 cm from Centrifuge must be kept and any breakable unit is moved away according to IEC61010 2 020 regulati...

Страница 8: ...o protect the internal chamber and erupting rotor Locking hook Peeping hole to check rpm by the digital speed meter Chamber Door packing Rotor Control board On off switch Emergency door open device ma...

Страница 9: ...mber to protect erupting rotor Motor cover intruding foreign substance into motor Motor shaft Don t grab the edge side of door to avoid any possible injury Put hands on the door cover and press and ch...

Страница 10: ...rotor tube adapter recommended by the manufacturer Check the proper tube rack insert and smooth motion Don t use the eroded motor of visibly mechanical failure 2 Tube arrangement Same volume of sample...

Страница 11: ...of the direction of the rotor 1 Weight loading on rotor Put tubes symmetrically on rotor only recommended tubes areworkable The weight difference between sample tubes must be minimized to diminish noi...

Страница 12: ...g bucket rotor Check the internal cleanness of bucket and wipe out the foreign material with dry fabric In the case of two buckets check the hanging status on rotor and do swing test manually Rotor mu...

Страница 13: ...same type symmetrically Use the adequate tubes with adapter Use always homogeneous tube same Weight material property density Volume 1 Caution 1 Balance loads and insert samples symmetrically Operate...

Страница 14: ...3 4 Function check Test maximum speed of rotor Check whether Centrifuge works at the door opened situation and door may open during operation 3 5 Operation detail 1 Control panel description...

Страница 15: ...2 Save and call program 3 Call input program 4 Input RPM RCF 5 Input Time 6 Short Run mode...

Страница 16: ...Requested Operation At DEC 0 an unit decelerates the same as speed as DEC 3 by 500rpm In less than 500rpm it stops without brakes...

Страница 17: ...7 Simulation Set value of simulation SL50 4 No 1 4 000 RPM 3 min is set in PROG No 3...

Страница 18: ...Exterior Clean the surface with soft dry fabric In the case of heavy contamination wipe out with soft soapy cloth and use dry cloth to remove water Don t use chemicals like alcohol benzene thinner whi...

Страница 19: ...urface treatment Acid or basic agent causes damage Imbalance of rotor weight by corrosion incurs heavy vibration noise eventually the physical damage Anti corrosive rotor may be affected by humidity s...

Страница 20: ...ed Rotor Check crack cleanness abrasion 3 RPM Install rotor and set RPM count the actual RPM with tachometer 4 Time Set 10 min and count the actual time with stop watch 5 Detecting Imbalance In the ca...

Страница 21: ...rheat 120 5 Temp hike inside of motor Check the stator of motor Switch power off 1 hour after re start E4 Low Speed RPM 10 Deficiency not reaching 200RPM within 2 sec Check and controller and motor Sw...

Страница 22: ...Specification Capacity Rotor Name Rotor No Max RPM Max RCF Max Radius mm Swing Rotor 85ml SL85 4 1 4 000 3 088 172 6 50ml SL50 6 2 4 000 3 088 172 6 Angle Rotor 15ml AL15 12 3 4 000 2 547 142 4 15ml...

Отзывы: