2. Installation
2-1. Safety
Centrifuge containing the high speed rotating part is dangerous. Follow
the instruction described in this manual to avoid breakage, trouble, injury.
1) Centrifuge must be placed the horizontally flat level, In the case of tilted
shaft of rotating big vibration occurs.
2) Check the voltage before connecting power, the improper volt may cause
the damage.
3) Use only the rotor, parts and accessories supplied by the maker, which is
not responsible for any damage caused by improper parts not
recommended by the maker.
4) Ascertain the material property completely and pre-arrange safety measure
in the case of handling pathogenic, toxic, radioactive material.
5) As Centrifuge is contaminated by pathogenic, toxic, radioactive material,
contaminant is to be cleared and ventilated and the isolation is needed.
6) Volatile and explosive material should not be treated by Centrifuge.
7) Don’t operate over the approved maximum rpm of rotor.
8) Don’t diminish and stop the rotating rotor with hand.
9) Rotor must be operated after symmetrical balance work of weight
10) Rotor must be fixed to the rotating shaft exactly and lid also fixed
completely. If not, the lid may open during operation.
11) Repairing and maintenance not mentioned on this manual must be
handled by the technician with the proper curriculum (in this case the
contaminant is to be cleared)
12) In order to keep the best condition and operation of long term the
chamber of rotor must be kept dry.
13) The chemical corrosion of rotor contacting strong acid, alkaline agent of
Cesium, silver, chlorine base should be avoided carefully.