background image

1. General Description

FLETA 40P is Multipurpose Centrifuge, this manual is for a starter, which

offers the information of trouble shooting, maintenance, precautionary

measure and usage.

Operator must read this manual for the successful handling of every

function and this manual should be placed at the easy accessible location.

All instructions described on this manual is for safe operation, therefore

You must read in advance to the actual operation.

1-1. Components

Item

Q’nty

Main-body                                                                                1

Power cable                                                                              1

User Manual                                                                             1

Hexagonal T wrench (5mm)                                                   1

Packing check list

1) Check thoroughly the status of the packing upon the arrival.

2) Immediately notify the detail in the case of the damage

3) The contact channel is on the packing and the user manual

1-2 The condition of keeping & operation

- Operation

Room temp 5~35

, max relative humidity 30~50%,

Air pressure 500~1,000  hpa

- Keeping

The ambient temp -10~+40

, max humidity 10~90%

Air pressure 500~1.060 hpa

Содержание FLETA 40P

Страница 1: ...Operation Manual FLETA 40P Multipurpose Centrifuge...

Страница 2: ...2 1 Safety 2 2 Location of Installation 2 3 Check list 2 4 Power Supply 3 Operation 3 1 Feature 3 2 Structure 3 3 Rotor 3 4 Function 3 5 Operating 4 Washing Maintenance 4 1 Washing 4 2 Maintenance 5...

Страница 3: ...s for safe operation therefore You must read in advance to the actual operation 1 1 Components Item Q nty Main body 1 Power cable 1 User Manual 1 Hexagonal T wrench 5mm 1 Packing check list 1 Check th...

Страница 4: ...minated by pathogenic toxic radioactive material contaminant is to be cleared and ventilated and the isolation is needed 6 Volatile and explosive material should not be treated by Centrifuge 7 Don t o...

Страница 5: ...ot be covered by the another instrument or fabric avoid the place of heavy dust Constant temp humidity The sensitive electronic controller is affected by ambient temp and humidity Avoid direct sunray...

Страница 6: ...d power supply to protect main body Door lock device Door sensor lets rotor stop in the case of open door and Centrifuge doesn t operate at the status of door open Auto door lock device doesn t allow...

Страница 7: ...ide of laboratory a remote area from Centrifuge to cut off power supply The safety distance 30 cm from Centrifuge must be kept and any breakable unit is moved away according to IEC61010 2 020 regulati...

Страница 8: ...o protect the internal chamber and erupting rotor Locking hook Peeping hole to check rpm by the digital speed meter Chamber Door packing Rotor Control board On off switch Emergency door open device ma...

Страница 9: ...mber to protect erupting rotor Motor cover intruding foreign substance into motor Motor shaft Don t grab the edge side of door to avoid any possible injury Put hands on the door cover and press and ch...

Страница 10: ...rotor tube adapter recommended by the manufacturer Check the proper tube rack insert and smooth motion Don t use the eroded motor of visibly mechanical failure 2 Tube arrangement Same volume of sample...

Страница 11: ...of the direction of the rotor 1 Weight loading on rotor Put tubes symmetrically on rotor only recommended tubes areworkable The weight difference between sample tubes must be minimized to diminish noi...

Страница 12: ...g bucket rotor Check the internal cleanness of bucket and wipe out the foreign material with dry fabric In the case of two buckets check the hanging status on rotor and do swing test manually Rotor mu...

Страница 13: ...same type symmetrically Use the adequate tubes with adapter Use always homogeneous tube same Weight material property density Volume 1 Caution 1 Balance loads and insert samples symmetrically Operate...

Страница 14: ...3 4 Function check Test maximum speed of rotor Check whether Centrifuge works at the door opened situation and door may open during operation 3 5 Operation detail 1 Control panel description...

Страница 15: ...2 Save and call program 3 Call input program 4 Input RPM RCF 5 Input Time 6 Short Run mode...

Страница 16: ...Requested Operation At DEC 0 an unit decelerates the same as speed as DEC 3 by 500rpm In less than 500rpm it stops without brakes...

Страница 17: ...7 Simulation Set value of simulation SL50 4 No 1 4 000 RPM 3 min is set in PROG No 3...

Страница 18: ...Exterior Clean the surface with soft dry fabric In the case of heavy contamination wipe out with soft soapy cloth and use dry cloth to remove water Don t use chemicals like alcohol benzene thinner whi...

Страница 19: ...urface treatment Acid or basic agent causes damage Imbalance of rotor weight by corrosion incurs heavy vibration noise eventually the physical damage Anti corrosive rotor may be affected by humidity s...

Страница 20: ...ed Rotor Check crack cleanness abrasion 3 RPM Install rotor and set RPM count the actual RPM with tachometer 4 Time Set 10 min and count the actual time with stop watch 5 Detecting Imbalance In the ca...

Страница 21: ...rheat 120 5 Temp hike inside of motor Check the stator of motor Switch power off 1 hour after re start E4 Low Speed RPM 10 Deficiency not reaching 200RPM within 2 sec Check and controller and motor Sw...

Страница 22: ...Specification Capacity Rotor Name Rotor No Max RPM Max RCF Max Radius mm Swing Rotor 85ml SL85 4 1 4 000 3 088 172 6 50ml SL50 6 2 4 000 3 088 172 6 Angle Rotor 15ml AL15 12 3 4 000 2 547 142 4 15ml...