2-4. Electrical requirements
Ascertain the rated voltage on the label of Centrifuge. Unclear power supply of
using should be checked again with the local electric power supplier or our A/S
center.
Warning ; Centrifuge must connect to ground.
- Centrifuge contains plug with earth pin which can be inserted into the grounded
receptacle. Using receptacle without earth must be replaced to grounded
receptacle, during this work user must be careful not to get electric shock with
the assistance of electrician. Avoid the receptacle without grounding.
- In the case of error the emergency switch is located at the exit door side or
outside of laboratory, a remote area from Centrifuge to cut off power supply.
- The safety distance 30 cm from Centrifuge must be kept and any breakable
unit is moved away according to IEC61010-2-020 regulation.
Electric safety
°
The power cord offered with Centrifuge is used only.
°
Connect power supply.
°
Don’t use the extended power code. Checking out grounded receptacle is
to be done by electrician.
°
Failure of earth work may cause the electric shock.
°
Another instrument should not step on the power code.
°
Ventilation hole should not be covered by any material.
°
Switch off power supply, pull off power code from receptacle, contact A/S
center in the below cases.
- Abnormal noise or smell
- Worn out or damaged power code
- Cut off wall mounted circuit breaker, fuse, another safety devices
- Spilt over liquid or water on Centrifuge
- Damaged part