QUICK SET UP AND CALIBRATION USING THE LOAD CELL mV/V SENSITIVITY
DAS72.1MKIII
QUICK SET UP AND CALIBRATION USING LOAD CELL mV/V (Continues on the next page)
Press the UP or DOWN key for more than 3 seconds to enter the Set-up Menu
Enter the
Setup menu
SET UP
3
4
Please note that if you wish to change any values in these sections, you have to first
press the recessed enable switch (top left corner of unit)
(Also see calibration examples on page 23)
Page 6
DAS72.1MKIII Manual v 4.xx
0
T
1
SET THE DISPLAY UNDER RANGE LIMIT. (MINIMUM VALUE -99,999)
.
Use this menu point to define the display minimum value. If the displayed
weight drops below this minimum, the display will then show under range
(dashes at the bottom of the display) This can be useful to indicate when a
weight value is outside a normal range. Use the UP/DOWN & MOVE RIGHT
keys to set the minimum display value and store by pressing the enter key.
Factory default value -9,999.
0
T
1
SET THE DISPLAY OVER RANGE LIMIT. (MAXIMUM VALUE +99,999)
Use this menu point to define the display maximum value. If the displayed weight
exceeds this maximum, the display will then show over range (dashes at the top of
the display). This can be useful to indicate when a weight value is outside a normal
range. Use the UP/DOWN & MOVE RIGHT keys to set the maximum display value
and store by pressing the enter key. Factory default value 10,000.
Set the Display
Upper Limit
Set the Display
Lower Limit
SELECT THE DECIMAL POINT POSITION ON THE DISPLAY (0, 0.0, 0.00,
0.000 or 0.0000)
Use the UP or DOWN key to set the required decimal point position on the
display. The factory default is no decimal place [0]. The maximum number of
places after the decimal point is four [0.0000].
Please note that setting a decimal place after defining the span calibration
value the will NOT increase the displayed resolution, it simply lights an LED on
the display in the appropriate position. For example if the span calibration value
is set to 1000 (no decimal point) and you then set one decimal place this will
mean a calibration value of 100.0. If you need the calibration value to be
1000.0 you will have to increase the span calibration value in menu 2.1 from
1000 to 10000
0
T
1
Set the Decimal
Point position