SET UP AND CALIBRATE ZERO FUNCTION
DATA COMMUNICATION SETTINGS (See Programmers Manual for full Communication Protocol)
0
T
1
3
4
0
T
1
0
T
1
T
T
T
2
3
4
SET THE BAUD RATE (9600, 19200, 38400, 57600, 115200). [FACTORY DEFAULT 9600]
Use the UP or DOWN key to set the required baud rate.
DAS72.1MKIII
SELECT EITHER RS422 OR RS485 INTERFACE (422, 485). [FACTORY DEFAULT RS485]
Use the UP or DOWN key to select either the RS422 or the RS485 interface.
0
T
1
T
T
T
2
0
T
1
3
4
SELECT AUTO TRANSMIT MODE (groS, nEt, ALL, SAP, AUEr, PEA, HOLd, UALL, PP, oFF).
Use the UP or DOWN key to select the auto transmit mode.:
groS
- the gross value is continuously
transmitted.:
nEt
- the net value is continuously transmitted.:
ALL
- the gross, net and status values are
continuously transmitted.
SAP
- the ADC value is continuously transmitted.:
AUEr
- the average value is
continuously transmitted.:
PEA
- the peak value is constantly transmitted.:
HOLd
- the hold value is
continuously transmitted.:
UALL
- the valley value is constantly transmitted.:
PP
- the peak to peak value is
constantly transmitted.:
oFF
- [FACTORY DEFAULT] the auto transmit is switched off.
0
T
1
T
T
T
2
0
T
1
SET THE DEVICE ADDRESS (0-255). [FACTORY DEFAULT 0]
Use the UP/DOWN & MOVE RIGHT key to set the device address (multi-drop) Set to 0 for single point to
point applications.
0
T
1
T
T
T
2
0
T
1
3
4
SET TRANSMIT DELAY (0-255 mS). [FACTORY DEFAULT 0]
Use the UP/DOWN & MOVE RIGHT key to set the Transmit Delay in milliseconds. This sets a delay time
after which the DAS 72.1 responds to a command. This is particularly handy in some PLC based systems
where the DAS72.1 response is too quick.
0
T
1
T
T
T
2
0
T
1
3
4
Page 22
DAS72.1MKIII Manual v 4.xx