LOGIC INPUTS  -  See Next Page

DAS72.1MKIII

ANALOGUE OUTPUT (4-20 mA) SETTINGS 

0

T

1

 3

 4

0

T

1

0

T

1

T

T

T

2

SET THE WEIGHT VALUE AT WHICH 4 mA IS SENT 

Use the UP/DOWN & MOVE RIGHT keys to set the weight value at which 4 mA is sent. This  typically will be set to 
zero, so that when the display reads zero weight, 4 mA will be sent down the analogue output. If you want the 
analogue output to run between 0 mA  and 20 mA, you will need to calculate what the 4 mA value will be and store 
it here.  To do this take the difference between the weight value at 20 mA and the weight value at 0 mA and divide it 
by the 20 and multiply by 4. For example if the analogue output is 0 - 20 mA  for 0 - 1000 kg then the 4 mA value will 
be 1000/20*4 = 200. 

 3

 4

0

T

1

T

T

T

2

TEST MODE FOR THE CURRENT OUTPUT (Range 0mA - 21mA)

Use the UP or DOWN & MOVE RIGHT key to set a current value in mA, which will be sent down the 
analogue output, independent of the load applied to the weighting system. If values below 0mA are selected 
the DAS will send its minimum value (0mA) until a value within its normal range is selected. Similarly if 
values above 21mA are selected then the DAS will send its maximum value (21mA) until a value within its 
normal range is selected.

0

T

1

SET THE WEIGHT VALUE AT WHICH 20 mA IS SENT

.

Use the UP/DOWN & MOVE RIGHT keys to set the weight value at which 20 mA is sent. Typically this will 
be set to the weight value at which you want 20 mA to be sent down the analogue output.

0

T

1

T

T

T

2

0

T

1

 3

 4

SET THE ANALOGUE OUTPUT BASE (groS, nEt [Default], PEA, AUEr, HoLd, PP, UALL, dISP, oFF)

Use the UP or DOWN key to set which parameter the analogue output will follow : 

groS

 - the analogue 

output follows the Gross value: 

nEt

 - the analogue output follows the Net value : 

PEA

 -  the analogue output 

follows the Peak value: 

AUEr

 - the analogue output follows the Average value: 

HOLd

 - the analogue output 

follows the Hold value: 

PP

 - the analogue output follows the Peak to Peak value: 

UALL

 - the analogue 

output follows the Valley value: 

dISP

 - the analogue output follows the value shown on the display: 

oFF

 - 

the analogue output is switched off.

0

T

1

T

T

T

2

0

T

1

 3

 4

Page 18

DAS72.1MKIII Manual v 4.xx

Содержание SENSOR SOEMER DAS 72.1 MKIII

Страница 1: ...45 4816 0880 Fax 45 4816 0870 www haubac com Tel 49 0 2721 9262 0 Fax 49 0 2721 9262 50 www soemer de Ideen Me technik Tel 44 0 1446 771185 Fax 44 0 1446 771186 www sensortechniques com TECHNIQUES LI...

Страница 2: ...Logic I O Status Page 14 4 Filter Setting and No Motion Page 16 Filter Setting Tables Update Rates Number of Readings Averaged Page 17 5 Analogue Output 4 20mA Settings Page 18 6 Logic Inputs 1 2 3 Pa...

Страница 3: ...out SET UP 1 Zero 1 0 allow 0 2 Calibrate 3 Set mV V 2 Span 1 Set cal n 2 Calibrate 3 Set mV V 4 Disp mV V 3 Display 1 o u limits n 2 Step n 3 Dec point 4 Logic stats 4 Filter 1 fcut Hz 2 Algorithm 3...

Страница 4: ...he display maximum value If the displayed weight exceeds this maximum the display will then show over range dashes at the top of the display This can be useful to indicate when a weight value is outsi...

Страница 5: ...1 kg steps with no decimal places then enter 3000 here 0 T 1 Calibrate the Span CALIBRATE THE SPAN CONVENTIONAL WEIGHING SYSTEM Use this menu point to calibrate the span using weights Apply the span...

Страница 6: ...int to define the display maximum value If the displayed weight exceeds this maximum the display will then show over range dashes at the top of the display This can be useful to indicate when a weight...

Страница 7: ...load cell stating amongst other things the Output at Rated Load ORL in mV V In single load cell applications the weighing system mV V value will be the load cell ORL value In applications using more...

Страница 8: ...Page 8 This page is intentionally left blank DAS72 1MKIII Manual v 4 xx...

Страница 9: ...value is less than the value set in Menu 1 4 5 factory default 2000 counts If you try to set a new zero which is greater than the value set in menu 1 4 5 you will see the error message Err 2 on the d...

Страница 10: ...1 T T T 2 0 T 1 0 T 1 T T T 2 ENABLE STORING OF TARE VALUE NON VOLATILE Use this menu point to enable t on the storing of the tare value in memory This will prevent the loss of the tare value in the e...

Страница 11: ...r key The default value is set to 2000 counts If the zero value you are trying to set is outside the range you will get the Err 2 message on the display Use the UP DOWN MOVE 0 T 1 0 T 1 T T T 2 ENABLE...

Страница 12: ...ll capacity here Use the UP DOWN MOVE RIGHT keys to enter the total load cell capacity For example if you have a weighing system consisting of 3 off 2000 kg load cells and you want to display in 1 kg...

Страница 13: ...ations the weighing system mV V value will be the load cell ORL value In applications using more than one load cell the weighing system mV V value will be the average of the individual load cells ORLs...

Страница 14: ...xample you select a display step size of 50 the display will show 000 050 100 150 etc as the weight value increases Using higher step sizes will give a more stable weight display but with a lower reso...

Страница 15: ...emporarily disabled whilst the logic input output status is shown momentarily The display then reverts back to displaying the weight value This can be useful when commissioning the system as you can s...

Страница 16: ...T 1 T T T 2 SELECT THE LOW PASS FILTER CUT OFF FREQUENCY RANGE 0 8 Use the UP DOWN keys to select the cut off frequency required A setting of 0 will disable the filter in both FIR IIR modes A setting...

Страница 17: ...in Stopband 90 90 90 90 90 90 90 90 Stopband in Hz 80 40 26 20 16 13 11 10 Update Rate in Samples sec 600 300 200 150 120 100 85 7 75 MENU 4 1 4 2 FILTER TYPE FIR MODE 1 COMBINATION LOW PASS Update Ra...

Страница 18: ...e selected the DAS will send its minimum value 0mA until a value within its normal range is selected Similarly if values above 21mA are selected then the DAS will send its maximum value 21mA until a v...

Страница 19: ...he status of Logic Input 1 If the input is high the display reads In1 1 If the input is low the display reads In1 0 In order for the status of this logic input to be read correctly you must set the fu...

Страница 20: ...ads In3 0 In order for the status of this logic input to be read correctly you must set the function as defined in section 6 3 1 to 0 No function T T T 2 0 T 1 T T T 2 T T T 2 NOT IN USE 3 4 3 4 3 4 3...

Страница 21: ...to define the parameter based on which Logic Output 1 will operate groS operates when the Gross value reaches or exceeds the setpoint nEt operates when the net value reaches or exceeds the setpoint P...

Страница 22: ...tinuously transmitted SAP the ADC value is continuously transmitted AUEr the average value is continuously transmitted PEA the peak value is constantly transmitted HOLd the hold value is continuously...

Страница 23: ...nabled See menu 1 1 DAS72 1MKIII ERROR CODES If you need to return the DAS72 1 to the factory settings simply press and hold the recessed enable switch whilst powering up the unit All parameters will...

Страница 24: ...ay to read the value of the calibration weight s which will be applied For this example if the calibration load applied which will be applied is 2000 kg set the display to read 2000 Press the 0 Enter...

Страница 25: ...value at which the sensitivity is known For this example the load cell capacity is 3 x 1000Kg 3000Kg 2mV V Set the display to read 3000 Press the 0 Enter key This defines the theoretical calibration w...

Страница 26: ...f the tank Therefore when there is 4000Kg in Example A rectangular tank fitted with 1 off 2000Kg 2mV V load cell central on the left hand side and a pivot hinge along the complete opposite face Dead l...

Страница 27: ...external up to 600 measurements per second FIR Filter 2 5 19 7 Hz or IIR Filter 0 25 18 Hz each programmable in 8 steps Software calibration and set up RS485 or RS422 full duplex 9600 115200 Baud up...

Отзывы: