Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
MEASure:DIODe?
Configures the instrument for diode tests. The displayed value will be reset and the instrument immediately triggers a
measurement.
Return values:
<Value>
The command returns a single reading.
Range: 5V
Usage:
Query only
MEASure:FREQuency[:VOLTAGE]? [{
<
Range
>
|AUTO|MIN|MAX|DEF}]
Configures the instrument for frequency measurements with main function AC V. The displayed values (including stati-
stic values) will be reset and the instrument immediately triggers a measurement. Display of frequency is only available
as 2nd function for the main functions AC V and AC I.
Return values:
<Range>
AC voltage: 400mV, 4V, 40V, 400V, 750V (5Hz to 700kHz)
AUTO:
Auto range
MIN: 400mV
MAX: 750V
DEFault: 400mV
If the input signal is greater than can be measured on the selected range (manual ranging),
the instrument returns 9.9037.
Usage:
Query only
MEASure:FREQuency:CURRent [{
<
Range
>
|AUTO|MIN|MAX|DEF}]
Configures the instrument for frequency measurements with main function AC I. The displayed values (including stati-
stic values) will be reset and the instrument immediately triggers a measurement. Display of frequency is only available
as 2nd function for the main functions AC V and AC I.
Return values:
<Range>
AC current: 20mA, 200mA (5Hz to 10kHz)
2A, 10A (5Hz to 5kHz)
AUTO:
Auto range
MIN: 20mA
MAX: 10A
DEFault: 20mA
If the input signal is greater than can be measured on the selected range (manual ranging),
the instrument returns 9.9037.
Usage:
Query only