Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
Bit No. Meaning
0
Operation Complete
This bit is set on receipt of the command *OPC exactly, when all previous commands have been
executed.
1
Not used
2
Query Error
This bit is set, if either the controller wants to read data from the instrument without having sent a
query, or if it does not fetch requested data and sends new instructions to the instrument instead.
The cause is often a query which is faulty and hence cannot be executed.
3
Device-dependent Error
This bit is set, if a device-dependent error occurs. An error message with a number between -300
and -399 or a positive error number is entered into the error queue.
4
Execution Error
This bit is set if a received command is syntactically correct, but cannot be performed for other rea-
sons. An error message with a number between -200 and -300 is entered into the error queue.
5
Command Error
This bit is set, if a command is received, which is undefined or syntactically incorrect. An error mes-
sage with a number between -100 and -200 is entered into the error queue.
6
Not used
7
Power On (supply voltage on)
This bit is set, when switching on the instrument.
Table 1.9: Bits of the event status register
(please refer to page 20)
STATus:OPERation Register
In the
CONDition
part, the register contains information which operations the instrument is being executing. In the
EVENt part, it contains information which operations the instrument has executed since the last reading. It can be read
using the commands
STATus:OPERation:CONDition?
or
STATus:OPERation[:EVENt]?.
The remote commands for the STATus:OPERation register are described on page 65.
Bit No.
Meaning
0
Calibrating
(for service department only)
1 to 3
Not used
4
Measuring
The bit is set, while the instrument is measuring.
5
Waiting for Trig
This bit is set while the instrument is waiting for the trigger.
6 to 9
Not used
10
Instrument Locked (RWLock)
11 to 15
Not used
Table 1.10: Bits of the STATus:OPERation register
(please refer to page 20)