Subject to change without notice
15
Amplitude Flatness: (sine/triangle)
0.05 Hz to 0.5 MHz:
max. 0.2 dB
0.5 MHz to 10 MHz:
max. 2.0 dB
DC-Offset:
variable (on/off, except impulse function)
Offset range:
into 50 Ω
max. ± 2.5V
open circuit
max. ± 5V
Trigger Output:
square wave synchronous
to output of approx.
+5V/TTL
FM Input
(VCF, BNC-connector on rear panel of HM8001-2 and
Opt. HO801)
Frequency change:
approx. 1 : 100
Input impedance:
6 kΩ II 25pF
Protection voltage:
± 30V max.
Internal Sweep
Sweep speed:
20 ms to 15 s
Sweep range:
approx. 1 : 100
General Information
Operating temperature:
+5°C . . . +40°C
Storage temperature:
–20°C . . . +70°C
Max. relative humidity:
5% . . . 80%
(without condensation)
Supply (from HM8001-2):
+5V/200mA
+16V/300mA;
-16V/250mA
(
∑
= 9.8 W).
Dimensions (without 22-pin flat connector)
W x H x D:
135 x 68 x 228 mm;
Weight:
approx. 800g
Operating Modes
Sine - Square - Triangle - DC - Pulse
free running, internal sweep, or external frequency modulati-
on, with or without DC Offset
Frequency Range
Total Range:
0.05 Hz . . . 10 MHz
(8 Decade Steps)
Variable Frequency
Adjustment:
x0.09 to x1.1 (12:1)
Frequency Stability:
<
0.5%/h or 0.8%/day
at constant ambient temperature
(medium frequency control position)
Waveform Characteristics
Sine Wave Distortion
0.05 Hz to 1 MHz:
max. 0.5%
1 MHz to 10 MHz:
max. 5%
Square Wave Risetime:
typ. 15 ns
Overshoot:
<
5%
(when output is terminated with 50 Ω)
Triangle Non-Linearity:
<
1% (up to 100 kHz)
Display
Frequency:
5 digit, 7 segment LED;
8 x 5mm each
Accuracy:
up to 5 Hz:
±(3% + 3 digit)
5 Hz to 10 MHz:
±(5 x 10
-5
+ 1 digit)
LED-Indicator for:
mHz, Hz, kHz and sec
Outputs
Signal Output:
short-circuit proof
Impedance:
50 Ω
Output Voltage
into 50 Ω
10 V
pp
open circuit
20 V
pp
Pulse Output Voltage
into 50 Ω :
5 Vpp
open circuit:
10 Vpp
Attenuation:
max. 60 dB
2 steps:
20 dB ±0.2 dB each
variable:
0 to 20 dB
S p e c i f i c a t i o n s
Specifications
All Data valid at 23 °C after 30 min. warm-up period
Included in delivery:
Function generator HM8030-6, Operating
manual
Optional Accessories:
BNC test cable HZ33/HZ34
50 Ω Through termination HZ22
Silicone Test lead HZ10 S/R
w w w.hameg.com
Subject to change without notice · © HAMEG Instruments GmbH · DQS-zertifiziert: DIN EN ISO 9001:2000, Reg. Nr.: DE-071040 QM
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49(0)61828000 · Fax +49(0)6182 800100 · www.hameg.com · [email protected]
A Rohde & Schwarz Company