background image

21

Subject to change without notice

First Time Operation

made. Thereafter the instrument will revert to its last used
operating mode.

If after approx. 20 seconds no trace is visible, the

 AUTO SET

pushbutton should be pressed briefly. This selects the Yt
mode and medium trace and readout intensity (

please note

”AUTO SET”

). Adjust

 Y-POS.I 

and

 X-POS. 

controls to center

the baseline. Adjust

 INTENS. 

(intensity) and

 FOCUS 

controls

for medium brightness and optimum sharpness (input(s)
grounded) of the trace. The oscilloscope is now ready for use.

If the

 AUTO SET 

function was not used and only a spot

appears (CAUTION! CRT phosphor can be damaged), reduce
the intensity immediately and check that the XY mode is not
selected (XY not displayed in the readout).

To obtain the maximum life from the cathode-ray tube, the
minimum intensity setting necessary for the measurement in
hand and the ambient light conditions should be used.

Particular care is required when a single spot is displayed, as
a very high intensity setting may cause damage to the
fluorescent screen of the CRT. Switching the oscilloscope off
and on at short intervals stresses the cathode of the CRT and
should therefore be avoided.

The instrument is so designed that even incorrect operation
will not cause serious damage.

Trace Rotation TR

In spite of Mumetal-shielding of the CRT, effects of the
earth’s magnetic field on the horizontal trace position cannot
be completely avoided. This is dependent upon the orientation
of the oscilloscope on the place of work. A centered trace may
not align exactly with the horizontal center line of the graticule.
A few degrees of misalignment can be corrected by a
potentiometer accessible through an opening on the front
panel marked TR.

Probe compensation and use

To display an undistorted waveform on an oscilloscope, the
probe must be matched to the individual input impedance of
the vertical amplifier.

For this purpose a square wave signal with a very fast rise
time and minimum overshoot should be used, as the sinusoidal
contents cover a wide frequency range.

The built-in calibration generator provides a square wave
signal with a very fast risetime (<4ns), and switch-selectable
frequencies of approx. 1kHz and 1MHz from the output
socket below the CRT screen.

As the squarewave signals are used for probe compensation
adjustments, neither the frequency accuracy nor the pulse
duty factor are of importance and therefore not specified. The
output provides 0.2Vpp ±1% (tr <4ns) for 10:1 probes. When
the Y deflection coefficient is set to 5mV/div, the calibration
voltage corresponds to a vertical display of 4 divisions (10:1
probe).
The output socket has an internal diameter of 4.9mm to
accommodate the internationally accepted shielding tube
diameter of modern Probes and F-series slimline probes. Only
this type of construction ensures the extremely short ground
connections which are essential for an undistorted waveform
reproduction of non-sinusoidal high frequency signals.

Adjustment at 1kHz

The C-trimmer adjustment (low frequency) compensates the
capacitive loading on the oscilloscope input. By this adjustment,
the capacitive division assumes the same ratio as the ohmic
voltage divider to ensure the same division ratio for high and
low frequencies, as for DC. (For 1:1 probes or switchable
probes set to 1:1, this adjustment is neither required nor
possible). A baseline parallel to the horizontal graticule lines is
essential for accurate probe adjustments. (See also ”Trace
rotation TR”).

Connect the probes (Types

 HZ51

,

 52

,

 54

, or

 HZ36

) to the

 CH

input. Set the deflection coefficient to 5mV/div and the input

coupling to DC. The time deflection coefficient should be set
to 0.2ms/div. All deflection coefficients should be calibrated.
Plug the probe tip into the calibrator output socket.

Approximately 2 complete waveform periods are displayed
on the CRT screen. The compensation trimmer should be
adjusted. The location of the low frequency compensation
trimmer can be found in the probe information sheet. Adjust
the trimmer with the insulated screw driver provided, until the
tops of the square wave signal are exactly parallel to the
horizontal graticule lines (see 1kHz diagram). The signal height
should then be 4div ± 0.16div (= 4 % (oscilloscope 3% and
probe 1%). During this adjustment, the signal edges will
remain invisible.

Adjustment at 1MHz

Probes

 HZ51

,

 52 

and

 54 

can also be HF-compensated. They

incorporate resonance de-emphasing networks (R-trimmer in
conjunction with capacitor) which permit probe compensation
in the range of the upper frequency limit of the vertical
oscilloscope amplifier. Only this compensative adjustment
ensures optimum utilization of the full bandwidth, together
with constant group delay at the high frequency end, thereby
reducing characteristic transient distortion near the leading
edge (e.g. overshoot, rounding, ringing, holes or bumps) to an
absolute minimum.

Using the probes

 HZ51

,

 52 

and

 54

, the full bandwidth of the

oscilloscope can be utilized without risk of unwanted waveform
distortion.

Prerequisite for this HF compensation is a square wave
generator with fast risetime (typically 4ns), and low output
impedance (approx. 50

), providing 0.2V at a frequency of

approx. 1MHz. The calibrator output of this instrument meets
these requirements when the CAL. pushbutton is depressed.

Connect the probe to

 CH I 

input. Depress the

 CAL. 

pushbutton

for 1MHz. Operate the oscilloscope as described under 1kHz
but select for 0.2µs/div time deflection coefficient setting.
Insert the probe tip into the output socket. A waveform will
be displayed on the CRT screen, with leading and trailing
edges clearly visible. For the HF-adjustment now to be
performed, it will be necessary to observe the rising edge as
well as the upper left corner of the pulse top. The location of
the high frequency compensation trimmer(s) can also be
found in the probe information sheet. These R-trimmer(s)

Содержание HM404-2.02

Страница 1: ...Instruments HANDBUCH MANUAL MANUEL Oscilloscope HM404 2 02 ENGLISH...

Страница 2: ...MANUAL HANDBUCH MANUEL...

Страница 3: ...fference measurement in DUAL mode Yt 23 Phase difference measurement in DUAL mode 23 Measurement of an amplitude modulation 23 Triggering and time base 24 Automatic Peak value Triggering 24 Normal Tri...

Страница 4: ...ia the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the osci...

Страница 5: ...0cm internal graticule Acceleration voltage approx 2000V Trace rotation adjustable on front panel Z Input Intens modulation max 5V TTL Calibrator 0 2V 1 1kHz 1MHz tr 4ns Line voltage 100 240V AC 10 50...

Страница 6: ...The instrument has been designed for indoor use The permissible ambient temperature range during operation is 10 C 50 F 40 C 104 F It may occasionally be subjected to temperatures between 10 C 50 F an...

Страница 7: ...nts The oscilloscope can be operated in any position but the convection cooling must not be impaired The ventilation holes may not be covered For continuous operation the instrument should be used in...

Страница 8: ...upling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general e...

Страница 9: ...example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must be co...

Страница 10: ...lse and calculation is unnecessary Calculation of the example in the figure above results in a signal risetime tr 162 8 752 22 13 25ns The measurement of the rise or fall time is not limited to the tr...

Страница 11: ...intain the nominal output voltage independent of frequency only if their connection cable is terminated with the prescribed resistance Here it must be noted that the terminating resistor HZ22 will onl...

Страница 12: ...ive If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a horizontal position shift DC voltage measurement If no signal is applied at the INPUT CHI 26...

Страница 13: ...trigger mode is automatically activated or not depends on the trigger coupling setting TRIG MODE The way the trigger point symbol in the readout responds on different LEVEL control knob settings indic...

Страница 14: ...control knob function between attenuator and vernier variable The current setting is displayed by the VAR LED located above the knob After switching the VAR LED 15 on the deflection coefficient is st...

Страница 15: ...the readout instead of CHP In alternate trigger mode the trigger point symbol is switched off Alternate triggering is not available or automatically switched off under the following conditions ADD add...

Страница 16: ...switch when the VAR LED above it is not lit Then the time deflection coefficient can be set in a 1 2 5 sequence and the time base is calibrated Rotating anticlockwise increases the deflection coeffic...

Страница 17: ...ains calibrated until the vernier knob is operated The readout now indicates T instead of T Rotating further anticlockwise increases the time deflection coefficient uncalibrated until the maximum is r...

Страница 18: ...afety earth contact of the line mains plug The input impedance is approx 1M II 20pF TRIG EXT This BNC socket is the external trigger signal input if external triggering is selected Briefly pressing th...

Страница 19: ...upted dotted line indicates the inactive cursor V t Pressing and holding this pushbutton changes from voltage to time or frequency measurement and vice versa In XY mode the instrument is automatically...

Страница 20: ...s MISCELLANEOUS and FACTORY 1 2 1 MISCELLANEOUS contains 1 2 1 1 CONTROL BEEP ON OFF In OFF condition the acoustic signals actuated by the control limits are switched off Note The default setting is O...

Страница 21: ...xtremely short ground connections which are essential for an undistorted waveform reproduction of non sinusoidal high frequency signals Adjustment at 1kHz The C trimmer adjustment low frequency compen...

Страница 22: ...deflects the beam in vertical direction while the time base causes an X deflection from left to right at the same time Thereafter the beam becomes blanked and fly back occurs The following Yt operati...

Страница 23: ...ng or lagging phase angle In alternate triggering condition phase difference measurement is not possible For greatest accuracy adjust the time base for slightly over one period and approximately the s...

Страница 24: ...l triggering Automatic Peak Value Triggering Instrument specific information can be drawn from the items NM AT 10 and LEVEL 12 in the section Controls and Readout This trigger mode is automatically se...

Страница 25: ...f the trigger signal and the lowest frequency range DC In this coupling mode the trigger signal is coupled galvanically to the trigger unit if normal triggering NM is present Therefore there is no low...

Страница 26: ...connected to a BNC connector for scope input via a shielded cable Between cable and BNC center conductor a resistor of at least 100 should be series connected RF decoupling Often it is advisable to sh...

Страница 27: ...se the holdoff control should be reset into its calibration detent fully ccw otherwise the brightness of the display is reduced drastically The function is shown in the following figures Fig 1 shows a...

Страница 28: ...search operation Photo 3 MODE DEL DELAY TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Delay time 20ms Reducing the time coefficient increasing the time base speed now expands the signa...

Страница 29: ...to the lower half of the CRT The 1mV div and 2mV div deflection coefficient will not be selected by AUTO SET as the bandwidth is reduced on these settings Attention If a signal is applied with a pulse...

Страница 30: ...or under test With high values of resistance the slope will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can...

Страница 31: ...luding probes between oscilloscope and circuit under test Otherwise both COMPONENT TESTER leads are not isolated against the circuit under test In circuit tests are possible in many cases However they...

Страница 32: ...cable length must be less then 3 meters and must contain 9 screened lines connected 1 1 The oscilloscope RS232 connection 9 pole D SUB female is determined as follows Pin 2 Tx data data from oscillos...

Страница 33: ...33 Subject to change without notice Front Panel HM404 2...

Страница 34: ...Subject to change without notice 34...

Страница 35: ...MANUAL HANDBUCH MANUEL...

Страница 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Отзывы: