Controls and Readout

(30) INPUT CH II 

- BNC socket.

This

 BNC 

socket is the signal input for channel II. The

outer (ground) connection is galvanically connected to
the instrument ground and consequently to the safety
earth contact of the line/mains plug.

In 

XY

 mode, signals at this input are used for the 

Y

deflection.

(31) AC / DC 

- Pushbutton with two functions.

Input coupling:

Briefly pressing this pushbutton switches over from

 AC

(

symbol) to

 DC 

(= symbol) input coupling and vice versa.

The setting is displayed in the readout with the deflection
coefficient.

Probe factor:

Pressing and holding the pushbutton selects the indicated
deflection coefficient of channel II displayed in the
readout, between 1:1 and 10:1. In condition 10:1 the
probe factor is thus indicated by a probe symbol displayed
by the readout in front the channel information (e.g.

”probe symbol”, Y2...

). In the case of cursor voltage

measurement, the probe factor is automatically included.

Please note:
The symbol must not be activated unless a x10 (10:1)
attenuator probe is used.

(32) GD - INV. -  

Pushbutton with two functions.

GD:

Each time this pushbutton is pressed briefly, the input is
switched from active to inactive and vice versa. It is
displayed in the readout as an earth (ground) symbol
instead of the deflection coefficient and the

 ~ (AC) 

or

 =

(DC) 

symbol.

The

 GD 

setting disables the input signal, the

 AC/DC (31)

pushbutton and the

 VOLTS/DIV (20) 

knob. In automatic

trigger mode the undeflected trace is visible representing
the 0 Volt trace position.

The 

”0 Volt”

 position is indicated by a 

””

 symbol in the

readout. Please note 

Y-POS. II (9)

.

INV.:

Pressing and holding this pushbutton switches the
channel II invert function on or off. The invert ”on”
condition is indicated by the readout with a horizontal
bar above 

”Y2” 

(Yt mode). The invert function causes

the signal display of channel II to be inverted by 180°.

(

33) TRIG. EXT. / INPUT (Z) 

- BNC socket with two functions.

The outer (ground) connection is galvanically connected
to the instrument ground and consequently to the safety
earth contact of the line/mains plug. The input impedance
is approx. 1M

 II 20pF.

TRIG. EXT:

This

 BNC 

socket is the external trigger signal input, if

external triggering is selected. Briefly pressing the

 TRIG.

(18) 

pushbutton, until the

 TRIG. ”EXT” -LED (18) 

is lit,

switches the external trigger input active.

The trigger coupling depends on the

 TRIG. MODE (21)

setting.

Z- Input:

If neither

 COMPONENT TEST 

nor external trigger

coupling

 (TRIG. EXT.) 

is chosen, the socket is operative

as a 

(trace intensity modulation) input.

High TTL level (positive logic) effects blanking, low level
gives unblanking. No higher voltages than +5 Volt are
permitted.

Below the CRT there are the controls for the readout,
the component tester and the squarewave calibrator
with their outputs.

Please note:
The following description of the cursor related controls
assumes that the readout is visible and the component
tester is switched off.

(34) MENU

 – Pushbutton

Pressing and holding the pushbutton activates the display
of  the 

MAIN MENU

. It contains the submenus 

SETUP

and 

CALIBRATE

.

Once a menu is displayed, the following pushbuttons are
of importance:

1. 

SAVE

 and 

RECALL

 (7) pushbutton.

Briefly pressing selects the submenu or an item within
the submenu.

2. 

SAVE

 (7) pushbutton 

SET

 function.

Pressing and holding the 

SAVE

 (7) pushbutton calls (sets)

the menu  or the previously selected item. In those cases
where the item is marked with 

ON / OFF

 the setting

changes from 

ON

 to 

OFF

 or vice versa.

If 

“PRESS SECURITY KEY!”

 is displayed, this function is

available only for HAMEG authorized service workshops.
Press 

AUTOSET

 (3) to switch back to the menu.

3. 

AUTOSET

 (3) pushbutton.

Each time the 

AUTOSET

 pushbutton is pressed the menu

is switched back one step until 

MAIN MENU

 is displayed.

Then pressing the 

AUTOSET

 pushbutton again switches

the menu operation off and 

AUTOSET

 is set to the normal

function.

(35) ON/OFF - CHI/II  - 1/

∆∆∆∆∆

t

This pushbutton has several functions.

ON/OFF:

Pressing and holding the pushbutton switches both cursor
lines  on or off. As the cursor lines are part of the readout,
they are visible only if the readout is switched on. If the
cursor lines are switched off and MEAN VALUE 

ON

 is

activated (Menu: SETUP > MISCELLANEOUS) the
instrument is prepared for DC Mean Value display. Please

Содержание HM404-2.02

Страница 1: ...Instruments HANDBUCH MANUAL MANUEL Oscilloscope HM404 2 02 ENGLISH...

Страница 2: ...MANUAL HANDBUCH MANUEL...

Страница 3: ...fference measurement in DUAL mode Yt 23 Phase difference measurement in DUAL mode 23 Measurement of an amplitude modulation 23 Triggering and time base 24 Automatic Peak value Triggering 24 Normal Tri...

Страница 4: ...ia the device under test mains line supply test leads control cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the interior of the osci...

Страница 5: ...0cm internal graticule Acceleration voltage approx 2000V Trace rotation adjustable on front panel Z Input Intens modulation max 5V TTL Calibrator 0 2V 1 1kHz 1MHz tr 4ns Line voltage 100 240V AC 10 50...

Страница 6: ...The instrument has been designed for indoor use The permissible ambient temperature range during operation is 10 C 50 F 40 C 104 F It may occasionally be subjected to temperatures between 10 C 50 F an...

Страница 7: ...nts The oscilloscope can be operated in any position but the convection cooling must not be impaired The ventilation holes may not be covered For continuous operation the instrument should be used in...

Страница 8: ...upling The input coupling is selectable by the AC DC pushbutton The actual setting is displayed in the readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general e...

Страница 9: ...example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 22 68nF must be co...

Страница 10: ...lse and calculation is unnecessary Calculation of the example in the figure above results in a signal risetime tr 162 8 752 22 13 25ns The measurement of the rise or fall time is not limited to the tr...

Страница 11: ...intain the nominal output voltage independent of frequency only if their connection cable is terminated with the prescribed resistance Here it must be noted that the terminating resistor HZ22 will onl...

Страница 12: ...ive If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a horizontal position shift DC voltage measurement If no signal is applied at the INPUT CHI 26...

Страница 13: ...trigger mode is automatically activated or not depends on the trigger coupling setting TRIG MODE The way the trigger point symbol in the readout responds on different LEVEL control knob settings indic...

Страница 14: ...control knob function between attenuator and vernier variable The current setting is displayed by the VAR LED located above the knob After switching the VAR LED 15 on the deflection coefficient is st...

Страница 15: ...the readout instead of CHP In alternate trigger mode the trigger point symbol is switched off Alternate triggering is not available or automatically switched off under the following conditions ADD add...

Страница 16: ...switch when the VAR LED above it is not lit Then the time deflection coefficient can be set in a 1 2 5 sequence and the time base is calibrated Rotating anticlockwise increases the deflection coeffic...

Страница 17: ...ains calibrated until the vernier knob is operated The readout now indicates T instead of T Rotating further anticlockwise increases the time deflection coefficient uncalibrated until the maximum is r...

Страница 18: ...afety earth contact of the line mains plug The input impedance is approx 1M II 20pF TRIG EXT This BNC socket is the external trigger signal input if external triggering is selected Briefly pressing th...

Страница 19: ...upted dotted line indicates the inactive cursor V t Pressing and holding this pushbutton changes from voltage to time or frequency measurement and vice versa In XY mode the instrument is automatically...

Страница 20: ...s MISCELLANEOUS and FACTORY 1 2 1 MISCELLANEOUS contains 1 2 1 1 CONTROL BEEP ON OFF In OFF condition the acoustic signals actuated by the control limits are switched off Note The default setting is O...

Страница 21: ...xtremely short ground connections which are essential for an undistorted waveform reproduction of non sinusoidal high frequency signals Adjustment at 1kHz The C trimmer adjustment low frequency compen...

Страница 22: ...deflects the beam in vertical direction while the time base causes an X deflection from left to right at the same time Thereafter the beam becomes blanked and fly back occurs The following Yt operati...

Страница 23: ...ng or lagging phase angle In alternate triggering condition phase difference measurement is not possible For greatest accuracy adjust the time base for slightly over one period and approximately the s...

Страница 24: ...l triggering Automatic Peak Value Triggering Instrument specific information can be drawn from the items NM AT 10 and LEVEL 12 in the section Controls and Readout This trigger mode is automatically se...

Страница 25: ...f the trigger signal and the lowest frequency range DC In this coupling mode the trigger signal is coupled galvanically to the trigger unit if normal triggering NM is present Therefore there is no low...

Страница 26: ...connected to a BNC connector for scope input via a shielded cable Between cable and BNC center conductor a resistor of at least 100 should be series connected RF decoupling Often it is advisable to sh...

Страница 27: ...se the holdoff control should be reset into its calibration detent fully ccw otherwise the brightness of the display is reduced drastically The function is shown in the following figures Fig 1 shows a...

Страница 28: ...search operation Photo 3 MODE DEL DELAY TIME DIV 5ms div Trigger coupling TV F Trigger slope falling Delay time 20ms Reducing the time coefficient increasing the time base speed now expands the signa...

Страница 29: ...to the lower half of the CRT The 1mV div and 2mV div deflection coefficient will not be selected by AUTO SET as the bandwidth is reduced on these settings Attention If a signal is applied with a pulse...

Страница 30: ...or under test With high values of resistance the slope will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can...

Страница 31: ...luding probes between oscilloscope and circuit under test Otherwise both COMPONENT TESTER leads are not isolated against the circuit under test In circuit tests are possible in many cases However they...

Страница 32: ...cable length must be less then 3 meters and must contain 9 screened lines connected 1 1 The oscilloscope RS232 connection 9 pole D SUB female is determined as follows Pin 2 Tx data data from oscillos...

Страница 33: ...33 Subject to change without notice Front Panel HM404 2...

Страница 34: ...Subject to change without notice 34...

Страница 35: ...MANUAL HANDBUCH MANUEL...

Страница 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Отзывы: