C o n t r o l s a n d R e a d o u t
29.1.3
50Ω / 1 MΩ Input Impedance
This function allows you to select the input impedance of chan-
nel 2. For measurement of low and high frequencies, voltages
at sources with impedances deviating from 50Ω an impedance
setting of 1 MΩ is recommended, together with the use of a
frequency compensated divider probe for minimum capacitive
load.
If the source impedance is 50Ω and the voltage to be measured
is within the voltage limits a 50Ω cable and 50Ω impedance
setting should be used.
29.2 Ground (GND) On Off
The pushbutton will alternate between switching the amplifi er
input to the signal or to ground.
If the signal is disconnected respectively the amplifi er input
connected to ground the readout will show a ground symbol
behind the sensitivity indication. In automatic trigger mode the
trace will be visible in a reference position which can be used as
a 0 V ground reference. The readout will show a symbol ( ) for
0 V which will be close to the vertical centre line of the graticule,
it is the zero reference for any measurements.
Referred to the trace position 0 V a DC voltage may be measured
after the input was returned to the signal.
29.3 Invert On Off
The pushbutton will alternate between not inverted and inverted.
An inverted signal will be indicated in the readout by bar above
the channel symbol. The trigger signal taken from an input will
not be affected.
29.4 Variable On Off
If activated (“On” intensifi ed) the CH2 VAR pushbutton
29
will
be illuminated and the VOLTS/DIV VAR knob
13
will function as
variable which can change the sensitivity continuously between
1 mV/cm to
>
20 V/cm. If in uncalibrated position, the readout
will replace ”:“ by ”
>
“ (e.g. ”CH2
>
5mV“.) and this also in any
cursor measurements indicating that the sensitivity is unca-
librated.
Pressing the “Variable On Off” function key again, switches
“Variable” off and the VOLTS/DIV VAR knob
13
back to calibrated
1-2-5 sequence defl ection coeffi cient function.
29.5 Probe
The menu display depends on whether the probe has dividing
ratio identifi cation. The value (automatically or manually selec-
ted) will be automatically taken into account regarding the
sensitivity indicated and measurements.
29.5.1
If a HAMEG probe with automatic dividing ratio identi-
fi cation is used, the menu display shows in normal intensity
“Probe” and beneath the dividing ratio (e.g. “*10”).
29.5.2
If no probe or a probe without automatic dividing ratio
identifi cation is used, the menu displays “Probe” and beneath
the last manually selected dividing ratio: the INTENS knob
2
symbol. Pressing the allocated function key causes “Probe” to be
displayed intensifi ed and the FOCUS TRACE MENU pushbutton
illuminated. The INTENS knob
2
then can be used for manual
selecting the dividing ratio of the applied probe. If no probe is
applied “*1” should be selected.
30
INPUT CH1
(BNC connector)
This is the CH1 signal input connector. In Yt mode it is a Y input,
in XY-mode it is the X signal input. The connector housing is
connected to the instrument housing and thus to safety ground.
The ring around the connector is the probe identifi cation contact,
no voltage may be applied here.
31
INPUT CH2
(BNC connector)
This is the CH2 signal input connector. It is a Y input in Yt and XY
mode. The connector housing is connected to the instrument
housing and thus to safety ground. The ring around the connector is
the probe identifi cation contact, no voltage may be applied here.
32
AUX pushbutton
This pushbutton belongs to AUXILIARY INPUT
33
(additional
input). The menu called by this pushbutton will depend upon
the actual operating mode.
VAR
VAR
VAR
x10
FOCUS
TRACE
VOLTS / DIV
VAR
VOLTS / DIV
VAR
TIME / DIV
VAR
X-POS
INTENS
!
LEVEL A/B
MENU
AUX
AUXILIARY INPUT
HORIZONTAL
INPUTS
1M
Ω
II13pF
max
250 Vp
CAT I
!
CAT I
!
DELAY
TRACE
SEP
CH 1/2
CURSOR
SAVE/
RECALL
AUTOSET
SETTINGS
HELP
POSITION 1
POSITION 2
VERT/XY
HM2005-2
ANALOG
OSCILLOSCOPE
200 MHz
TRIGGER
MODE
FILTER
SOURCE
TRIG
’
d
NORM
HOLD OFF
AUTO
MEASURE
CURSOR
MEASURE
TRIG. EXT. / Z-INP.
CH 1
CH 2
X-INP
CH 1
CH 2
HOR
MAG
1M
Ω
II
13pF
max
100 Vp
5 V
1 mV
5V
1 mV
0.5s
20ns
CH I MENU
AC/DC/50
Ω
GND
50
Ω
/ 1 M
Ω
INVERT
ON / OFF
VARIABLE
ON / OFF
PROBE
1 : 1 / 10 / 100
CH I: 500 mV
POWER
POWER
PROBE
ADJ
A N A L O G S C O P E
Instruments
MENU
MENU
OFF
OFF
COMP.
TESTER
1
2
3
4
5
6
7
8
15
22
23
19
20
16
17
24
18
21
25
26
9
11
10
13
12
14
37
27
30
28
29 31
32
33
36
35
34