5
Subject to change without notice
S p e c i f i c a t i o n s
200 MHz Analog Oscilloscope HM2005-2
All data valid at 23 °C after 30 minute warm-up
Vertical Deflection
Channels:
2
Operating Modes:
CH 1 or CH 2 separate, DUAL (CH 1 and
CH 2 alternate or chopped), Addition
X in XY-Mode:
CH 1
Invert:
CH 1, CH 2
Bandwidth (-3 dB):
2 x 0…200 MHz
Rise time:
‹ 1,75 ns
Bandwidth Limiter
(switchable)
:
approx. 20 MHz (1 mV/div.…5 V/div.)
Deflection Coefficients (CH 1, 2):
12 calibrated steps
1…2 mV/div.:
± 3 % (0…100 MHz (-3 dB))
5 mV…5 V/div.:
± 3 % (1-2-5 sequence)
variable (uncalibrated):
› 1 mV/div.…5 V/div., continuous
Inputs CH 1, 2:
Impedance:
1 MΩ II 13 pF
Coupling:
DC, AC, 50 Ω, GND (ground)
Max. Input Voltage:
250 V (DC + peak AC), 50 Ω ‹ 5 V
rms
Y Delay Line (analog):
70 ns
Measuring Circuits:
Measuring Category I
Auxiliary input:
Function (selectable):
Ext. Trigger, Z (unblank)
Coupling (Ext. Trig./Z):
all / AC, DC
Max. input voltage:
100 V (DC + peak AC)
Triggering
Automatic (Peak to Peak):
Min. signal height:
5 mm
Frequency range:
10 Hz…250 MHz
Level control range:
from Peak- to Peak+
Normal (without peak):
Min. signal height:
5 mm
Frequency range:
0…250 MHz
Level control range:
-10…+10 div.
Operating modes:
Slope / Video
Slope:
Rising, falling, both
Sources:
CH 1, CH 2, alt. CH 1/2 (
≥
8 mm), Line, Ext.
Coupling:
AC:
10 Hz…250 MHz
DC:
0…250 MHz
HF:
30 kHz…250 MHz
LF:
0…5 kHz
Noise Rej. switchable
Video:
pos./neg. Sync. Impulse
Standards:
525 Line / 60 Hz Systems
625 Line / 50 Hz Systems
Field:
even/odd/both
Line:
all/line number selectable
Source:
CH 1, CH 2, Ext.
Indicator for trigger action:
LED
External Trigger via:
AUXILIARY INPUT (0.3 V
pp
, 0…200 MHz)
Coupling:
AC, DC
Max. input voltage:
100 V (DC + peak AC)
2nd Trigger
Min. signal height:
5 mm
Frequency range:
0…250 MHz
Coupling:
DC
Level control range:
-10…+10 div.
Horizontal Deflection
Time Base
Operating modes:
A, ALT (alternating A/B), B
Time base A:
20 ns/div.…0.5 s/div. (1-2-5 sequence)
Time base B:
20 ns/div.…20 ms/div. (1-2-5 sequence)
Accuracy A and B:
± 3 %
X Magnification x10:
to 2 ns/div.
Accuracy:
± 5 %
Variable time base A/B
: cont. 1:2.5
Hold Off time:
var. 1:10 (LED-Indication)
XY Mode
Bandwidth X-Amplifier:
0…3 MHz (-3 dB)
X Y phase shift:
‹ 3 ° ‹ 220 kHz
Accessories supplied:
Line Cord, Operators Manual and Software for Windows
on CD-ROM, 2 Probes 10:1 (HZ200), RS-232 Interface (HO710)
Optional accessories:
HO720
Dual-Interface USB / RS-232
HO730
Dual-Interface Ethernet / USB
HO740
Interface IEEE-488 (GPIB)
HZ13
Interface cable (USB) 1,8m
HZ14
Interface cable (serial) 1:1
HZ20
Adapter, BNC to 4mm banana
HZ33
Test cable 50Ω, BNC/BNC, 0,5m
HZ34
Test cable 50Ω, BNC/BNC, 1m
HZ45
19''-Rackmount Kit 4RU
HZ51
Probe 10:1 (150MHz)
HZ52
Probe 10:1 RF (250MHz)
HZ53
Probe 100:1 (100MHz)
HZ56-2 AC/DC Current probe
HZ70
Opto-Interface (with optical fiber cable)
HZ72
GPIB-Cable 2m
HZ100
Differential probe 20:1 / 200:1
HZ109
Differential probe 1:1 / 10:1
HZ115
Differential probe 100:1 / 1000:1
HZ154
Standard probe 1:1 / 10:1
HZ350
Probe 10:1 with automatically identification (350MHz)
HZ355
Slimline probe 10:1 with automatically identification (500MHz)
HZO20
High voltage probe 1000:1 (400MHz,1000Vrms)
HZO30
Active probe 1GHz (0,9pF, 1MΩ, including many accessories)
HZO50
AC/DC Current probe 20A, DC...100kHz
HZO51
AC/DC Current probe 1000A, DC...20kHz
Operation / Measuring / Interfaces
Operation:
Menu (multilingual), Autoset, Help functions (multilingual)
Save / Recall internal:
9 Instrument parameter settings
Signal sources:
CH 1, CH 2
Frequency counter:
6 digit resolution:
› 1 MHz…250 MHz
5 digit resolution:
0.5 Hz…1 MHz
Accuracy:
50 ppm
Auto Measurements:
Frequency, Period, V
dc
, V
pp
, V
p+
, V
p-
Cursor Measurements:
Δt, 1/Δt (f), tr, ΔV, V to GND, ratio X, ratio Y
Resolution Readout / Cursor:
1000 x 2000 Pts
Interfaces:
RS-232
1)
Display
CRT:
D14-375GH
Display area (with graticule):
8 div. x 10 div.
Acceleration voltage:
approx. 14 kV
General Information
Component tester
Test voltage:
approx. 7 V
rms
(open circuit), approx. 50 Hz
Test current:
max. 7 mA
rms
(short circuit)
Reference Potential:
Ground (safety earth)
Probe ADJ Output:
1 kHz /1 MHz square wave signal 0.2 V
pp
(tr ‹ 4 ns)
Trace rotation:
electronic
Line voltage:
105…253 V, 50/60 Hz ± 10 %, CAT II
Power consumption:
42 Watt at 230 V, 50 Hz
Protective system:
Safety class I (EN61010-1)
Operating temperature:
+5…+40 °C
Storage temperature:
-20…+70 °C
Rel. humidity:
5…80% (non condensing)
Dimensions
(W x H x D)
:
285 x 125 x 380 mm
Weight:
5.6 kg
1)