background image

12

Subject to change without notice

20 V

pp

) HAMEG offers the external 50 Ω termination HZ22. For 

probes, terminations are neither required nor allowed, they 
would ruin the signal.  

Probes feature very low loads at fairly low frequencies: 10 MΩ 
in parallel to a few pF, valid up to several hundred kHz. How-
ever, the input impedance diminishes with rising frequency to 
quite low values. This has to be borne in mind as probes are, 
e.g., entirely unsuitable to measure signals across high impe-
dance high frequency circuits such as bandfi lters etc.! Here 
only FET probes can be used. Use of a probe as a rule will also 
protect the scope input due to the high probe series resistance 
(9 MΩ). As probes cannot be calibrated precisely enough during 
manufacturing individual calibration with the scope input used 
is mandatory! (See Probe Calibration).

Passive probes will, as a rule, decrease the scope bandwidth and 
increase the rise time. We recommend to use HZ200 probes in 
order to make maximum use of the combined bandwidth. HZ200 
features 2 additional hf compensation adjustments. 

Whenever the DC content is 

>

 250 V

DC

 coupling must be used in 

order to prevent overstressing the scope input capacitor. This is 
especially important if a 100:1 probe is used as this is specifi ed 
for 1200 V

DC

 + peak AC. 

AC coupling of low frequency signals may produce tilt.

If the DC content of a signal must be blocked, it is possible to 
insert a capacitor of proper size and voltage rating in front of the 
probe, a typical application would be a ripple measurement. 

When measuring small voltages the selection of the ground 
connection is of vital importance. It should be as close to voltage 
take off point as possible, otherwise ground currents may de-
teriorate the measurement. The ground connections of probes 
are especially critical, they should be as short as possible and 
of large surface area (litz wire; skin effect). 

STOP

 If a probe is to be connected to a BNC connector, 

use a probe tip to BNC adapter.

If ripple or other interference is visible, especially at high sen-
sitivity, one possible reason may be multiple grounding. The 
scope itself and most other equipment are connected to safety 
ground, so ground loops may exist. Also, most instruments will 
have capacitors between line and safety ground installed, which 
conduct current from the live wire into the safety ground. 

First time operation and initial adjustments  

Prior to fi rst time operation the connection between the instru-
ment and safety ground must be ensured, hence the plug must 
be inserted fi rst.

Use the red POWER pushbutton to turn the scope on. Several 
displays will light up. The scope will then assume the set up, 
which was selected before it was turned off. If no trace and 
no readout are visible after approximately 20 sec, push the 
AUTOSET button.

As soon as the trace becomes visible select an average inten-
sity with INTENS, then select FOCUS and adjust it, then select 
TRACE ROTATION and adjust for a horizontal trace.

With respect to crt life, use only as much intensity as necessary 
and convenient under given ambient light conditions. When not 
in use, turn the intensity fully off rather than switching the scope 
on and off too much as this is detrimental to the life of the crt 
heater. Do not allow a stationary point on the screen, it might 
burn the crt phosphor.

With unknown signals start with the lowest sensitivity 5 V/cm, 
connect the input cables to the scope, and then to the measuring 
object which should be de energized beforehand. Then turn the 
measuring object on. If the trace disappears, push AUTOSET.

Trace rotation TR

The crt has an internal graticule. In order to adjust the defl ected 
beam with respect to this graticule the Trace Rotation control 
is provided. Select the function Trace Rotation and adjust for a 
trace which is exactly parallel to the graticule.

Probe adjustment and use

In order to ensure proper matching of the probe used to the 
scope input impedance the oscilloscope contains a calibrator 
with short rise time and an amplitude of 0.2 V

pp

 ±1 %, equivalent 

to 4 cm at 5 mV/cm when using 10:1 probes.

The inner diameter of the calibrator connector is  4.9 mm and 
standardized for series F probes. Using this special connec-
tor is the only way to connect a probe to a fast signal source 
minimizing signal and ground lead lengths and to ensure true 
displays of pulse signals.

1 kHz – adjustment

This basic adjustment will ensure that the capacitive attenuation 
equals the resistive attenuation thus rendering the attenuation 
of the probe independent of frequency. 1:1 probes can not be 
adjusted and need no such adjustment anyway. 
Prior to adjustment make sure that the trace rotation adjust-
ment has been performed. 

F i r s t   t i m e   o p e r a t i o n   a n d   i n i t i a l   a d j u s t m e n t s          

 incorrect 

correct 

incorrect

Содержание HM2005-2

Страница 1: ...2 0 0 M H z A n a l o g O s c i l l o s c o p e H M 2 0 0 5 2 Manual English...

Страница 2: ...s the oscilloscope may be effected by such elds Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT gap As the bandwidth of each ampli er stage...

Страница 3: ...12 Trace rotation TR 12 Probe adjustment and use 12 1 kHz adjustment 12 1 MHz adjustment 13 Operating modes of the vertical ampli er 13 XY operation 14 Phase measurements with Lissajous gures 14 Measu...

Страница 4: ...5 60 and 625 50 standard 250MHz 6 Digit Frequency Counter Cursor and Automatic Measurement 14kV high writing speed CRT Readout Autoset Delay Line no Fan Save Recall Memories for Instrument Settings He...

Страница 5: ...Hz 3dB X Y phase shift 3 220kHz Accessories supplied Line Cord Operators Manual and Software for Windows on CD ROM 2 Probes 10 1 HZ200 RS 232 Interface HO710 Optional accessories HO720 Dual Interface...

Страница 6: ...in into the next locking position Handle mounting dismounting The handle can be removed by pulling it out further depending on the instrument model in position B or F Safety The instrument ful ls the...

Страница 7: ...tions are valid after a 30 minute warm up period between 15 and 30 degr C Speci cations without tolerances are average values Warranty and repair HAMEG instruments are subjected to a strict quality co...

Страница 8: ...n 28 Calls selectable trigger lter coupling noise reject and trigger slope menu 18 SOURCE pushbutton 29 Calls trigger source menu e g CH1 CH2 Alt 1 2 External AC Line 19 TRIG d LED 29 Lit on condition...

Страница 9: ...in the menu hierarchy F r o n t P a n e l E l e m e n t s B r i e f D e s c r i p t i o n VAR VAR VAR x10 FOCUS TRACE VOLTS DIV VAR VOLTS DIV VAR TIME DIV VAR X POS INTENS LEVEL A B MENU AUX AUXILIAR...

Страница 10: ...which coupling was chosen stands for DC stands for AC For hf measurement an in ternal 50 terminator can be activated which is indicated by an symbol in the readout Amplitude of signals In contrast to...

Страница 11: ...e and probe may be neglected For the measurement of rise times proceed mainly as outlined above however rise times may be measured anywhere on the screen It is mandatory that the rising portion of the...

Страница 12: ...ent from the live wire into the safety ground First time operation and initial adjustments Prior to rst time operation the connection between the instru ment and safety ground must be ensured hence th...

Страница 13: ...following rule Always look at the two signals in the one channel only or the dual modes not in ADD mode and make sure that they are within the permissible input signal range this is the case if they...

Страница 14: ...ollowing formulas and a pocket calculator with trigonometric functions This calculation is independent of the signal amplitudes Please note As the trigonometric functions are periodic limit the cal cu...

Страница 15: ...degree will result a b a b m or m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the amplitude and time variables can be used without any in uence on the result Triggering and...

Страница 16: ...h will cause a trigger irrespective of the polarity of the next slope Rising slope means that a signal comes from a negative po tential and rises towards a positive one This is independent of the vert...

Страница 17: ...f possible the coil should be shielded without creating a shorted winding Alternate trigger This mode is selected with SOURCE 18 Alt 1 2 The readout will display Tr alt but no trigger point symbol ind...

Страница 18: ...ne period of a signal is displayed at a convenient time base setting Increasing the sweep speed with TIME DIV will expand the display from the start so that parts of the signal will disappear from the...

Страница 19: ...disconnected from safety ground Except for the two test leads there may be no further connection between scope and component See Tests within a circuit As described in section Safety all ground conne...

Страница 20: ...d of FET With depletion type MOSFETs and all JFETs the channel will conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very low it may look like a...

Страница 21: ...ith the interface HO710 the scope rmware can be updated via the interface The RS 232 interface has the usual 9 pole D Sub connector The scope can be remotely controlled via this bidirectional interfac...

Страница 22: ...tely upon leaving the menu Each menu is assisted by HELP texts which can be called by pressing the HELP 8 and which will be also displayed by the readout If HELP was called and the INTENS knob 2 moved...

Страница 23: ...layed pressing the CURSOR MEASURE pushbutton again then causes the display of the Cursors menu and its selection box Additionally the FOCUS TRACE MENU pushbutton 3 will light indicating that the INTEN...

Страница 24: ...rmine for which signal channel the CURSOR measurements are valid This is done by showing the INTENS knob signal next to the channel number After selection of the signal source the CURSOR lines must th...

Страница 25: ...l 2 POSITION 2 will function as Y position control of CH2 in Yt mode and if the CH1 2 CURSOR TRACE SEP pushbutton 11 is not illuminated 6 SETTINGS pushbutton Pressing this pushbutton will open the SET...

Страница 26: ...atic probe factor or manual probe setting dependent The readout will display the calibrated sensitivity e g CH1 5mV Depending on the sensitivity selection the signal will be shown with smaller or grea...

Страница 27: ...pe Thereafter the LEVEL A B control will control the time base B trigger a second trigger point symbol will be shown and marked with B turning it CW will increase the sensitivity The sensitivities can...

Страница 28: ...led as XY displays can not be triggered 17 1 Menu Slope The menu Edge will appear if Edge was selected in the Trig ger menu to be called with MODE 16 pushbutton and after the FILTER 17 pushbutton was...

Страница 29: ...ing dark X position control green Delay time control 22 1 X POS If the pushbutton doesn t lit the HORIZONTAL knob 23 functions In BOTH MODE both slopes will trigger This allows e g the display of eye...

Страница 30: ...selected and the option B variable On Off was set to Off Turning the control CCW will decrease turning it CW will in crease the time base speed The speed can be selected between 20 ms cm 50 ns cm in a...

Страница 31: ...se A Further information can be found in the section HOLD OFF adjustment in the chapter Triggering and time bases 25 MAG x10 pushbutton Pressing this pushbutton will turn on the x 10 magni er No menu...

Страница 32: ...be determined with respect to the formerly set zero reference 27 3 Invert On Off unavailable in XY mode This pushbutton will alternate between not inverted or inverted of the CH1 signal The readout wi...

Страница 33: ...inputs and not from the added signal there is no true reference for the trigger point symbol the symbol will thus be switched off However the LEVEL A B control 15 is active 28 5 XY mode In this mode C...

Страница 34: ...ation The value automatically or manually selec ted will be automatically taken into account regarding the sensitivity indicated and measurements 29 5 1 If a HAMEG probe with automatic dividing ratio...

Страница 35: ...tment of 10 1 probes The frequency can be selected by pressing the pushbutton PROBE ADJ and calling the menu Utilities Further information may be found in the section Probe adjust ment and use in the...

Страница 36: ...hwarz Company D 63533 Mainhausen registered trademark Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de w w w h a m e g c o m 41 2005 02...

Отзывы: