1
Critical Note on Calibration
When configuring the R-1307 Readout, it is critical to match the target ID with the target ID
identified on the serial number of the target. For example, if the target ID on the target serial
number label is 2, then the R-1307 must also be set to the number 2. If the target and readout and
not matched, a centering error of up to .002 in. (0.05 mm) can occur. In addition, the laser switch
setting (CONT.
or Fixed
vs. PULSE) must also agree with the R-1307 Readout setting. For
CONTinuous (Fixed) Mode, set the readout to F10.10 and for Pulsed Mode, set the R-1307 to
P10.10.
For example:
.
.
or
.
.
for R-1307 #2
For more information on the Pulse/CONTinuous modes on the laser, see
Pulse/Continuous Modes (L-
705, L-706 and L-708 Lasers)
on Page 6. For complete information on matching the target to the readout,
see
Configuring the R-1307 or R-1307C for a Local Target
on Page 16.