Step 2: Target Mounting Error
a.
Insert the A-512 into the A-514 Adapter and
insert them into the
Near Bore
, making sure
the adapter's level vial is level. Then adjust
the L-102 to zero the display and center the
laser into the first reference bore (the graphic
below shows how the L-102 moves the height
of the laser beam, but not the angle).
b.
Press
Record
to record data for the target in the NORMal position. Rotate the target 180 degrees
(INverted position) and reinsert into the
Near Bore
. Press
Record
to record data for the target in the
INverted position. Rotate target 180 degrees again back to the NORMal position and reinsert into the
Near Bore
. The TSCE Mounting Error Offset is calculated and applied to each target reading. This
removes any remaining centering errors in the target and adapter.
Note:
Step 2: Target Mounting Error of the Bore9 program may be skipped if measuring bore
straightness or alignment tolerances are greater than .0005 in. (0.013 mm).