28
Step 8: Move the Target Back to the
Near Bore
Remove the target from the
Far
reference bore and
reinsert it into the
Near Bore
. Be sure to place the target
in the
exact
location that it was in previously, with the
cable at 6 o’clock. Plug the target into the readout and
power on the readout. Using the coarse adjustment of the
L-102 beam translator, adjust the laser beam until the
readout displays less than .0015 in. both vertically and
horizontally.
Remote Buck-in Note:
If the distance from the laser to the
target in the Near Bore is more than 10 percent of the distance
from the target in the Near Bore to the target in the Far Bore, it
will be necessary to follow the Remote Buck-In procedure
(see
Appendix C on Page 42).