35
■
Removing the water circulation hose
Turn off the LCOS-SLM and peripherals (Include a circulating water controller).
Check that the circulating water has stopped.
Remove the circulating water hose on the circulating water controller side.
Air out the water in the circulating water hose and heat sink.
・
Blow air into one side of the hose. Install a water absorption sheet (Wastes, towels, etc.) and a saucer
on the outlet side of the hose to prevent splashing of water.
・
Please blow air until water stops coming out.
Places the LCOS-SLM head in a stable plane.
Remove the hoses from the LCOS-SLM head one at a time to wipe off the water.
・
Take care not to splash water, and pull out the hose while pressing the stop ring against the main body.
Caution
If the medicine is not to be used for a long period of time, remove
water and store in a completely dried condition.
■
Attaching the Rod Stand to the Optical Adjustment Stage
The water-cooled type has a special hole (M6) for mounting the rod stand as the normal type.
In addition, it can be fixed to the optical stage of 50 mm pitch or 32 mm pitch using the mounting hole of the
bottom.
Fig.25
water-cooled L-shaped adapter drawing