34
■
Starting and circulating water
・
Check that the circulating water is flowing normally.
・
Ensure that the circulating water flow rate is 0.5 L/min. If the flow rate and water pressure cannot be ensured,
replace the water circulating hose immediately.
・
Do not stop the circulating water during laser irradiation.
Caution
Do not stop water circulation during high-power laser irradiation.
The device may break due to a sudden rise in temperature.
■
Hose for circulating water
A hose with an outside diameter of φ6mm can be used.
SMC water hose (For example, TU0604 series ) is recommended.
■
Installing a water circulation hose
Check that the circulating water controller is stopped and that the power of the LCOS-SLM is OFF.
Places the LCOS-SLM head in a stable plane.
Insert the circulating water hose into the WATER connector.
・
Insert the hose into the WATER connector (Direction of the photo arrow).
・
Insert the hose securely into the WATER connector until it stops.
Connect the circulating water hose to the circulating water controller.
Turn on the circulating water controller and check that there is no water leakage in addition to the
circulating water circulating normally.
Fig.24
Installing the circulating water hose
stopping
Water Connector
Hose for
circulating water