@X4-0047-5
- 18 -
5.4.
Operation by the remote
(
Command Set
)
When creating your own program by utilizing control commands, the following command
is available. In all commands, it is necessary to be terminated by CR(Carriage Return).
Refer to the control command list shown below.
Success
Failure
Set the high voltage output
V<value1><value2><CR>
value1
:
0h
~
14h
value2
:
0h
~
63h
VA
BC,BA
Set the maximum voltage
M<value1><value2><CR>
value1
:
0h
~
14h
value2
:
0h
~
63h
VA
BC,BA
Set the step voltage
S<value1><value2><CR>
value1
:
0h
~
14h
value2
:
0h
~
63h
VA
BC,BA
Set LCD Contrast
C<value><CR>
value
:
0h
~
9h
VA
BC,BA
Set the high voltage output ON/OFF
O<value><CR>
value
:
0h or 1h
VA
BC,BA
Set LCD backlight ON/OFF
B<value><CR>
value
:
0h or 1h
VA
BC,BA
Set the high voltage output switch LED ON/OFF
L<value><CR>
value
:
0h or 1h
VA
BC,BA
Set device mode : remote or local
Q<value><CR>
value
:
0h or 1h
VA
BC,BA
Read the device setting
R<CR>
None
13 byte data
BC
Read the high voltage monitor value
T<CR>
None
2 byte data
BC
Read the low voltage output (±5V
,
±15V)
A<CR>
None
8 byte data
BC
Reset the device to the condition at power-on.
D<CR>
None
VA
BC
Read Information
F<CR>
None
24 byte data
BC
Operation
Command
Argument
Response
The communication setting of RS-232C is as follows:
Baud rate : 9600 bit per second
Data length : 8 bits
Parity bit : None
Stop bit : 1 bit
Flow control : None
The details of each command are described from the next page.