(3) 4×4 binning readout
Exposure and 4
4 binning readout repeat according to the sequence set internally.
Command
: AMD N
(Free running mode)
Scan mode
: SMD S
(Binning)
Exposure time : NMD F, FBL n (Frame blanking)
Binning
: SPX 4
(4×4 binning)
1V (vertical)
: 28.0 ms
Setting count
: 1V to 36V
(28.0 ms to 1.00 s
)
Repeat cycle
: 35.7 Hz/n
Exposure time can be calculated using the formula below.
Exposure time = n × 28.0 ms
(n=1 to 36)
(4) Sub-array readout
Readout of the sub-array repeats along with frame blanking according to the sequence
set internally with the selection of <
SVO n/SVW m/SPX e
>. The exposure time
(electronic shutter time) is set with the
FBL n
command.
Command
: AMD N
(Free running mode)
Scan mode
: SMD A
(Sub-array)
Exposure time : NMD F, FBL n (Frame blanking)
Sub-array
: SVO n / SVW m / SPX e
Exposure time = FBL n
CCD exposure
Readout signal
from CCD chip
Digital data
output
Readout
period
Exposure time = FBL n
CCD exposure
Readout signal
from CCD chip
Digital data
output
Readout
period