qCMOS Camera C15550-20UP Instruction manual_Ver.1.0
66
*
5 At full resolution reatout.
*
6 Valid 4 digits and rounded up to 5th digit.
*
7 At Lightsheet readout mode.
*
8 At Area readout mode and Light sheet readout mode.
[Frame rate] *
1
at CoaXPress
at USB 3.1 Gen 1
at Full resolution
Standard scan
120 fps
17.6 fps
Ultra quiet scan
5.00 fps
5.00 fps
at Vertical 4 lines
Standard scan*
2
19 841 fps
4084 fps
Ultra quiet scan
826 fps
826 fps
*
1 Valid 3 digits and rounded down to 4th digit.
*
2 When using frame bundle function on DCAM-API.
(3) Mechanical specifications
Cooling method
Forced-air cooled / Water cooled
Cooling temperature
Forced-air cooled
(Ambient temperature: +25 °C)
-20 °C
Water cooled
(Water temperature: +25 °C, Ambient temperature: +25 °C)
-20 °C
Water cooled (Max cooling)
(Water temperature: +20 °C, Ambient temperature: +20 °C)
Less than
-35 °C *
1
Lens Mount
C-mount
*
1 Typical value.
*
2 The water temperature is +20 °C and the ambient temperature is +20 °C
(4) Power supply specifications
Camera
Input power supply
DC 12 V
Typical output
---
Power consumption
84 W
AC adapter
Input power supply
AC 100 V to AC 240 V
50 Hz / 60 Hz 2.5 A
Typical output
DC 12 V
8.34 A
Power consumption
155 VA
Note
•
Fluctuations of input power supply voltages are not to exceed ±10% of the nominal voltage.
(5) Operating environment
Ambient operating temperature
0 ºC to + 40 ºC
Ambient storage temperature
-10 ºC to + 50 ºC
Ambient operating humidity
30 % to 80 %, no condensation
Ambient storage humidity
Less than 90 %, no condensation
Place of operating
Indoor, altitude up to 2000 m
(6) Dimensional outline and weight
Dimensional outline
110 mm
(W)
× 118 mm
(H)
× 141.6 mm
(D)
Weight
Approx. 3.0 kg (Camera only)
•
Be careful not to drop off the camera or not drop underfoot when making it move because it
is approx. 3.0 kg.
Note
•
Refer to 15 “DIMENSIONAL OUTLINES” for detail of dimensions.
(7) Applicable standards
EMC
EN61326-1: 2013 Class A
FCC
47 CFR FCC Part15 Subpart B Class A