C11440-22CU/C11440-22CU01 Instruction manual_Ver.1.4
(2) USB 3.0
Vn
= Number of vertical line (The center of the set area is the middle of the sensor.)
Exp1 = 1.004 ms to 10 s
Exp2 = 1.05 ms to 10 s
1H
= 9.74436 × 10
-6
int () = The decimal point is rounded down.
•
The Exp1 and Exp2 values must be input to the calculation formula below in units
of seconds.
[Calculation formula]
Binning
Horizontal
Vertical
Calculation formula
Free running mode
1×1
1024
―
1/ (int(Vn/2048/30/1H)×1H)
1536
2048
1×1
512
―
1/ (int(Vn/2048/100/1H)×1H)
2×2
―
―
4×4
―
―
External trigger mode
(Edge trigger)
1×1
1024
112
≦
Vn
≦
2048 1/ (int(Vn/2048/30/1H)×1H)
1536
2048
1×1
1024
8
≦
Vn
≦
104
1/ (Vn/2×1H+1H×9+Exp1)
1536
2048
1×1
512
―
2×2
―
―
4×4
―
―
External trigger mode
(Level trigger)
1×1
1024
112
≦
Vn
≦
2048 1/ (int(Vn/2048/30/1H)×1H)
1536
2048
1×1
1024
8
≦
Vn
≦
104
1/ (Vn/2×1H+1H×4+Exp2)
1536
2048
1×1
512
―
2×2
―
―
4×4
―
―
External trigger mode
(Synchronous trigger
readout)
1×1
1024
24
≦
Vn
≦
2048
1/ (int(Vn/2048/30/1H)×1H)
1536
2048
1×1
64
8
≦
Vn
≦
16
1/(Vn/2×1H+1H×18)
1024
1536
2048
1×1
512
―
2×2
―
―
4×4
―
―
28
Содержание C11440-22CU
Страница 2: ...Blank page...
Страница 45: ...C11440 22CU C11440 22CU01 Instruction manual_Ver 1 4 Level trigger mode Figure 10 15 43...
Страница 46: ...C11440 22CU C11440 22CU01 Instruction manual_Ver 1 4 44...
Страница 50: ...C11440 22CU C11440 22CU01 Instruction manual_Ver 1 4 Start trigger mode Figure 10 22 Ex rising edge 48...
Страница 69: ...C11440 22CU C11440 22CU01 Instruction manual_Ver 1 4 15 DIMENSIONAL OUTLINES 15 1 C11440 22CU Unit mm 67...
Страница 70: ...C11440 22CU C11440 22CU01 Instruction manual_Ver 1 4 15 2 C11440 22CU01 Unit mm 68...