C11440-22CU/C11440-22CU01 Instruction manual_Ver.1.4
10-2-7 SLOW SCAN
It can change from the standard scan (Default) to the slow scan with a command.
Read-out noise is reduced with slow scan.
Note
•
Please refer to Figure 10-2-2 FRAME RATE CALCULATION about the frame rate value and
calculation formula with slow scan.
10-2-7-1 Timing Diagram
The read-out time is 33 ms at full frame and the frame rate is 30 frames/s at fastest.
The timing diagram for slow scan is similar to the timing diagram for the standard scan except
readout time. Please refer to the timing diagram of standard scan for each trigger mode.
Figure 10-13
41
Содержание C11440-22CU
Страница 2: ...Blank page...
Страница 45: ...C11440 22CU C11440 22CU01 Instruction manual_Ver 1 4 Level trigger mode Figure 10 15 43...
Страница 46: ...C11440 22CU C11440 22CU01 Instruction manual_Ver 1 4 44...
Страница 50: ...C11440 22CU C11440 22CU01 Instruction manual_Ver 1 4 Start trigger mode Figure 10 22 Ex rising edge 48...
Страница 69: ...C11440 22CU C11440 22CU01 Instruction manual_Ver 1 4 15 DIMENSIONAL OUTLINES 15 1 C11440 22CU Unit mm 67...
Страница 70: ...C11440 22CU C11440 22CU01 Instruction manual_Ver 1 4 15 2 C11440 22CU01 Unit mm 68...