10-3-5 PROGRAMMABLE TIMING OUTPUT
By using the programmable timing output, synchronizing with external devices is simple. A
system which needs simple timing signal does not require a delay unit or pulse generator. It is
possible to program and output a pulse that has an optional pulse width and an optional delay
time to the end of readout timing, Vsync or Hsync. The range of delay is 0 µs to 10 s, and the
range of pulse width is 10 µs to 10 s.
The relation between the parameter which can be set with each reference signal, and an output
signal becomes as shown below.
Reference signal
Output signal
Read End
Camera outputs a pulse after certain delay from the end of sensor readout. Also
the pulse width can be set.
Vsync
Camera outputs a pulse after certain delay from the beginning of readout. Also
the pulse width can be set.
Hsync
Camera outputs a pulse after certain delay from the ending of readout. Also the
pulse width can be set.
Figure 10-23 Programmable timing output