C10910 INSTRUCTION MANUAL Ver.1.19E
65
.
(3) EXT.GATE
This mode is usable with any combination regardless of what plug-in module is
connected. In this mode, the streak tube operates only during any desired time period.
This operation is shown in the timing chart below.
The gate opens on the rising edge of a pulse input via the GATE TRIG IN connector,
and closes on the falling edge. Input the pulses at the timing that matches the type of
measurement.
During this mode, only the MCP gate operates and the photocathode gate remains
open.
10-6 DELAY
The delay can be set on synchroscan units such as M10911-01, M10911-03, M10915-01,
etc. Delay periods of one cycle or more can be set with a plug-in unit alone. The control
value range is from 0 to 65 535.
10-7 TRIGGER MODE
"TRIG. MODE" can be selected only with single-sweep units (M10912-01, M10913-01,
M10913-11, M10916-01, M10916-02, M10916-03 etc). There are two settings available
with the Trigger Mode: "Continuous mode" and "Single mode".
10-7-1 CONTINUOUS MODE
This mode allows automatically accepting the next sweep trigger whenever the next sweep
can start after finishing the sweep from the previously received sweep trigger. Continuous
operation can be performed by sweep triggers only.
Gate status
Sweep trigger
Sweep waveform
Monitor Out
OPEN
CLOSE
About 300 ns(gate opening time after gate trigger input)
Gate trigger
> 300 ns (input after gate opened)
_
Содержание C10910
Страница 2: ......
Страница 89: ...C10910 INSTRUCTION MANUAL Ver 1 19E 87 Spacer ring for output optics ...
Страница 90: ...C10910 INSTRUCTION MANUAL Ver 1 19E 88 16 2 C10910 POWER SUPPLY UNIT Weight approx 5 9 kg Unit mm ...
Страница 92: ...C10910 INSTRUCTION MANUAL Ver 1 19E 90 16 4 M10911 02 SYNCHROSCAN UNIT Weight approx 3 85 kg Unit mm ...
Страница 94: ...C10910 INSTRUCTION MANUAL Ver 1 19E 92 16 7 M10913 11 SLOW SINGLE SWEEP UNIT Weight approx 3 4 kg Unit mm ...
Страница 95: ...C10910 INSTRUCTION MANUAL Ver 1 19E 93 16 8 M10914 01 HORIZONTAL BLANKING UNIT Weight approx 3 0 kg Unit mm ...
Страница 96: ...C10910 INSTRUCTION MANUAL Ver 1 19E 94 16 9 M10914 02 SHIFT BLANKING UNIT Weight approx 2 2 kg Unit mm operate ...
Страница 97: ...C10910 INSTRUCTION MANUAL Ver 1 19E 95 16 10 M10915 01 SYNCHRONOUS BLANKING UNIT Weight approx 2 85 kg Unit mm ...
Страница 99: ...C10910 INSTRUCTION MANUAL Ver 1 19E 97 16 12 A1976 01 INPUT OPTICS Weight approx 1 3 kg Unit mm ...
Страница 100: ...C10910 INSTRUCTION MANUAL Ver 1 19E 98 16 13 A1974 INPUT OPTICS Weight approx 2 1 Kg Unit mm depth 3 ...
Страница 102: ...C10910 INSTRUCTION MANUAL Ver 1 19E 100 16 16 A11771 01 MOUNT TABLE FOR ORCA R2 Weight approx 690 g Unit mm ...
Страница 103: ...C10910 INSTRUCTION MANUAL Ver 1 19E 101 16 17 A11771 03 MOUNT TABLE FOR ORCA FLASH4 0 Weight approx 1 0 kg Unit mm ...