6.0
Mounting the Electromagnetic Locks
6.1
Description
CAUTIONS
•
When the BloodTrack
®
Kiosk V5 relay door lock interface is used to drive a
magnetic lock, a diode must be fitted as per
Section 8.3.7
, otherwise the
BloodTrack Kiosk V5 may suffer from unpredictable behavior.
•
Use only the Haemonetics-approved Xtra-Guard
®
cabling type noted below.
•
Use
only
the Haemonetics-approved EcoWire
™ protective earth cable noted
below.
Access to the storage location is controlled by the BloodTrack Courier
®
software
driving the BloodTrack Kiosk V5 relay outputs. These relay outputs are used in
conjunction with electromagnetic locks installed on the blood product storage
location using various brackets.
) encased in a plastic enclosure (BTR4 or
BTR5 fridge door lock magnet) is optionally supplied with the BloodTrack Kiosk V5.
The signals are carried on this cable type which connects to the plug in terminal
block which interfaces to the BloodTrack Kiosk V5:
•
Alpha Wire EcoWire Part No. 6718 GY005.
The safety protective earth wire can also be seen. The protective earth cable used
with the door lock magnet must
only
be the following Haemonetics-approved cable
type:
•
Alpha Wire EcoWire 12 AWG E163869-DB RU AWM 11028 105C 600V
VW-1---CRU AWM I A/B 105C 600V FT1.