BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
4
•
For US:
•
FCC CFR 47 Part 15B:2010 for US.
2.1.3 Safety Standards
The following safety standards have been applied to the BloodTrack Kiosk V5:
•
IEC 60950-1:2006 2nd edition
•
IEC 60601-1:1995 2nd edition Safety Requirements For Medical Electrical Systems
•
IEC 60601-1:2005 3rd edition Safety Requirements For Medical Electrical Systems
•
CAN/CSA-C22.2 No. 60601-1:14.
The BloodTrack Kiosk V5 has been tested to these national variants of IEC 60950,
Information technology equipment
– Safety:
•
UL 60950-1:2007/R:2014-10
•
CAN/CSA C22.2 No. 60950-1:2007/A2:2014-10
•
EN 60950-1/A12:2011.
2.2
Test Summary
The BloodTrack Kiosk V5 was tested to:
•
EN 60601-1-2 medical device standard for EMC emission and immunity
•
EN55022 and EN 55024 emission and immunity standards for information
technology equipment
•
EN 61326 EMC emission and immunity for electrical equipment for
measurement, control and laboratory use.