BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
35
•
The BloodTrack Kiosk V5 should not have other equipment placed on top of
it as this could impede air flow for cooling.
When choosing a wall for mounting the BloodTrack Kiosk V5, be sure to consider
cable lengths and limitations, as well as wall structure.
Before beginning this procedure, verify that all parts were received
—see
Section 4.2.1
Notes:
•
If mounting the scanning shelf, it is advised that the active shelf be
assembled prior to mounting the BloodTrack Kiosk V5
—see
Section 5.1
•
All necessary checks should be made before drilling for fixings
—for
example, hidden wiring.
To mount the BloodTrack Kiosk V5 on a wall:
1. It is recommended that the BloodTrack Kiosk V5 be mounted with the lowest
wall mounting hole at a distance of 45.3 in. (1150 mm) above the floor
—see
. This is the best position to accommodate users of varying heights to
make use of the touch-screen and barcode scanner.
Ensure hardware is affixed to a wall or plywood mounted on a wall.
2. Ensure adequate clearance for the safe opening and closing of the storage
location door.
3. The wall mount requires a minimum of four secure fixings to the wall
—see
Содержание BloodTrack Kiosk V5
Страница 2: ...2016 Haemonetics Corporation...