BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
13
It is recommended that the PAT be performed on the BloodTrack Kiosk V5 at
regular intervals. Usually the requirement for PAT will be dependent upon on the
healthcare site’s health and safety policy for fixed equipment, but the time period for
the PAT should be determined in compliance with applicable health and safety
legislation, and by a risk analysis, to take into account the environment in which the
BloodTrack kiosk V5 is being used.
Operators of the BloodTrack Kiosk V5 should be aware of the requirements of
Section 3.0
Upon request, Haemonetics may provide system diagrams, schematics, component
part lists, descriptions, calibration instructions or other information that will assist
with any servicing or repairs made to the BloodTrack Kiosk V5.
2.13 Cleaning Specification
WARNING
!
Never spray water or any liquid substance directly on any part of the exterior of
the BloodTrack
®
Kiosk V5 as the electric insulation could become seriously
damaged and cause a malfunction. Spray any cleaning fluid onto the cloth first,
and then apply to the unit.
CAUTION
Do not touch, push or rub the touch-screen with glass, tweezers or anything
harder than a finger.
The BloodTrack Kiosk V5 will need occasional cleaning. In particular, the touch-
screen on the BloodTrack Kiosk V5 will need cleaning to remove finger marks and
dust. Wipe off any moisture drops as soon as possible as their long-time contact
with the touch-screen will cause deformations and color-fading.
It is recommended that the exterior of the device be cleaned monthly or earlier if
obviously dirty.
Ambersil
®
Amberclens
®
Anti-Static Foaming Cleaner is recommended to clean the
touch-screens, using a lint-free clean soft cloth. Amberclens is safe for use on the
touch-screen, as well as the exterior surfaces of the BloodTrack Kiosk V5.
Amberclens is safe for use on all metals and plastics, if rinsed after application. After
cleaning, wipe the surface with a clean cloth dampened with water.
Alternatively, a neutral detergent or isopropyl alcohol may be used.
Do
NOT
use any kind of chemical solvent, acidic or alkali solution.
Do
NOT
use cleaning sprays on the touch-screen as many cleaning solutions
contain abrasive elements
—such as Acetone or Toluene—that will cause chemical
damage to the touch-screen.
Содержание BloodTrack Kiosk V5
Страница 2: ...2016 Haemonetics Corporation...