BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
15
3.0
Safety Precautions
Follow all cautions and instructions marked on the equipment and listed in this
guide. Keep this guide available for future reference.
Observe the following safety precautions when setting up equipment:
WARNINGS
•
To avoid risk of electric shock, this equipment must only be connected
to supply mains with protective earth.
•
To avoid risk of electric shock during servicing/maintenance, the
power to this equipment must be disconnected when accessing
internal components
.
•
The BloodTrack
®
Kiosk V5 is shipped with a grounding type (three-wire)
power cord. To reduce the risk of electric shock, the cord
MUST
be plugged
into a grounded power outlet that provides a protective earth.
•
The BloodTrack Kiosk V5 is designed to work with single-phase power
system having a grounded neutral conductor. To reduce the risk of electric
shock, do not plug the BloodTrack Kiosk V5 into any other type of power
system. Contact a facilities manager or a qualified electrician if unsure what
type of power is supplied to the building.
•
RF energy is used by the BloodTrack Kiosk V5 only for its internal function.
The RF emissions of the BloodTrack Kiosk V5 are very low and are not likely
to cause any interference in nearby electronic equipment. The BloodTrack
Kiosk V5 should not be operated nearby other equipment susceptible to
interference. See
Appendix B
for further guidance and advice on ways to
avoid or minimize such interference.
•
The BloodTrack Kiosk V5 should not be used adjacent to or stacked with
other equipment. In case adjacent or stacked use is necessary, the
BloodTrack Kiosk V5 should be observed to verify normal operation in the
configuration in which it will be used.
•
Do not operate the BloodTrack Kiosk V5 with the outer plastic cover
removed. Failure to take this precaution may result in personal injury and
system damage.
•
Do not simultaneously touch a patient and any accessible contacts of
connectors, contacts of fuse holders or accessible parts exposed by removal
of a cover on the BloodTrack Kiosk V5.
•
Never push objects of any kind through openings in the equipment.
Dangerous voltages may be present. Conductive foreign objects could