Amplification adjustment with calibration weights
 

(3. Menu step)

 

- adjust zero point (2. menu step)

 

- apply calibration weights to measuring a roll

 

- select the menu point "Cal" in the adjustment mode

 

- enter the percentage value of the calibration weight*

 

- store adjustment

increase

decrease

adjustment range:

10.0 …110.0 %

Fault conditions (Display)

"Gain":  The calculated amplification is out of range and is displayed in menu step 4.  

 

 

(calculated amplification as gain value)

"null":   The calibration weight is too small.

*Model the web geometry with a cable or rope and apply calibration weight. The weight should approxi-

mate the actual web tension in the operating mode. The DMA2 offers the facility to increase the weight with 

a percentage calculation. As an example: 70% of the required 100 kg are applied to the measuring roll.  

In the adjustment mode "Cal" 70.0 are entered representing the percentage value.

Caution: If during zero point adjustment a value unequal to 0 has been entered then value adjustments are 

possible only in steps of + 10%.

The step velocity 

increases with 

the key holding 

time

measuring roll

calibration weight

amplifier

rope or 

cable

08.08 DE1DM2EA.indd 

Содержание DMA2

Страница 1: ...in 2 Amplification adjustment with calibration weights 2 Connection Wiring 3 Operation Preparation 4 Adjustment mode 4 Pre Adjustments Operationg Modes Display intensity 4 Filter type of the current output 4 Output signal range 5 Display value 5 Adjustement Of The Parameters Automatic zero point adjustment 6 Manual zero point adjustment 6 Entering amplification as gain value 6 Amplification adjust...

Страница 2: ...inal rating 1 5 m V V bridge supply voltage 10 V the output signal is therefore 15 mV 0 015 V at nominal force Amplifier output 10 V at nominal web tension 0 V at zero web tension If the sensor is operating below its nominal force then the amplification must be increased Example The maximum sensor load is only 75 of nominal force The output signal is therefore only this leads to Amplification adju...

Страница 3: ...y from the shield with PE Additional recommendation Connect the 0 V of the 24 V circuitry at a defined place prefably in close proximity of the PE If this reference point is to be operated in the floating point mode then the voltage between 0 V and PE should not exceed 50 Vpp The four terminal blocks with five terminals each can be separatly plugged in simplifying trouble shooting in case of fault...

Страница 4: ...re and the LED lights up until the procedure is completed The amplifier switches now automatically into the operating mode Leaving the adjustment mode without storage When exceeding the waiting time of 20 seconds the amplifier switches without changes automatically into the operating mode An alternative is pressing the menu key beyond the 8 menu step Display intensity 7 Menu step select the menu s...

Страница 5: ...ds to 10 V output voltage increasing The various decimal ranges are stepped through sequentially The step velocity increases with the key holding time This adjustment deletes stored peak values Output signal range 8 Menu step select the menu point output signal range in the adjustment mode select operating mode store adjustment Operating Mode Signal value Voltage output Current output Option C Opt...

Страница 6: ...sing adjustment range 10 0 100 Entering amplification as gain value 4 Menu step calculate amplification siehe section Adjustment Modes on page 2 if no sensor is connected then terminals 18 19 or 13 14 have to be connected with each other Select the menu point Gain in the adjustment mode select the desired amplification store the adjustment increasing decreasing adjustment range 400 2800 page 6 See...

Страница 7: ...nd is displayed in menu step 4 calculated amplification as gain value null The calibration weight is too small Model the web geometry with a cable or rope and apply calibration weight The weight should approxi mate the actual web tension in the operating mode The DMA2 offers the facility to increase the weight with a percentage calculation As an example 70 of the required 100 kg are applied to the...

Страница 8: ...the SET key max value min value Fault display during operation Exceeding the maximum output signal voltage of approximately 13 3 V e g resulting from a broken sensor cable HAEHNE Elektronische Messgeräte GmbH Heinrich Hertz Str 29 D 40699 Erkrath Germany Telefon 0211 9 2591 0 Fax 0211 9 2591 20 http www haehne de Email info haehne de HAEHNE page 8 08 08 DE1DM2EA indd ...

Отзывы: