Display value
(5. Menu step)
select the menu point "display value" in the adjustment mode.
- selection alternatives:
value of the numerical display in % (factory adjustment: 100 without a decimal point)
or
as actual value in the corresponding dimensional unit (decimal point display)
- store adjustment
100 % corresponds to
10 V output voltage
increasing
The various decimal ranges are
stepped through sequentially.
The step velocity increases with the
key holding time
This adjustment deletes stored peak values!
Output signal range
(8. Menu step)
select the menu point "output signal range"
in the adjustment mode
-select operating mode
- store adjustment
Operating Mode
Signal
value
Voltage
output
Current output
Option C
Option N
bipolar
(factory
adjustment)
-100 %
-10 V
4 mA
0 mA
0 %
0 V
4 mA
0 mA
+ 100 %
+10 V
20 mA
20 mA
unipolar
-100 %
0 V
4 mA
0 mA
0 %
+5 V
12 mA
10 mA
+100 %
+10 V
20 mA
20 mA
The operating modes "bipolar" or
"unipolar" must be selected before zero
point and amplification adjustment!