![Hadeco Bidop ES-100V3 Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/hadeco/bidop-es-100v3/bidop-es-100v3_operating-manual_577369056.webp)
49
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Not applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The ES-100V3 is intended for use in the electromagnetic environment specified below. The customer
or the user of the ES-100V3 should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge(ESD)
IEC61000-4-2
±8kV contact
±
2kV, ±4kV, ±8kV,
±15kV air
±8kV contact
±
2kV, ±4kV,
±8kV, ±15kV air
Floors should be wood,
concrete or ceramic tile. If floors
are converted with synthetic
material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC61000-4-4
±2kV for power supply
lines(100KHz)
±1kV for input/output
lines
Not applicable
Surge
IEC61000-4-5
±1kV differential mode
±2kV common mode
Not applicable
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
Dip to 0% for 1 cycles
@ 0°phase angle
Dip to 70% for 25/30
cycles @ 0°phase
angle
Dropout to 0% for
0.5cycle @ 0°, 45°, 90°,
135°, 180°, 225°,
270°&315°
Interrupts
100% for 250/300
cycles
Not applicable
Power frequency
(50Hz) magnetic
field
IEC61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test revel.
Guidance and manufacturer’s declaration – electromagnetic immunity
The ES-100V3 is intended for use in the electromagnetic environment specified below. The customer or
the user of the ES-100V3 should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Compliance
level
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the ES-100V3, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d
= 1,2√
P