![Hadeco Bidop ES-100V3 Скачать руководство пользователя страница 54](http://html.mh-extra.com/html/hadeco/bidop-es-100v3/bidop-es-100v3_operating-manual_577369054.webp)
47
8-3. Specifications
Probes:
Model
Freq.
Ispta* (in situ) [mW/cm
2
]
BT2M20S8C
2 MHz
80 or less
BT4M05S8C
4 MHz
390 or less
BT5M05S8C
5 MHz
390 or less
BT8M05S8C
8 MHz
390 or less
BT10M5S8C 10 MHz
390 or less
* Ispta: Special Peak-Temporal Average Intensity.
Battery:
DC 9 volts, Alkaline square type battery
Battery life:
Approx. 2.5 hours (When using with alkaline battery and
backlight off.)
Automatic shut-off No signal: 2 min.
Freeze: 5 min.
Others: 15 min. (only FHR WAVE mode: 35 min.)
Frequency range:
80 / 200 Hz to 5 kHz
Mode settings:
Memory, Waveform, Direction, Time scale, Others
Waveform memory: 30 waveforms
LCD display:
128 x 64 dots, STN LCD with backlight
Bi-directional waveform (normal & slow mode)
Numerical data (Systolic, diastolic & mean velocities, RP,
PI, SD, HR)
Heart rate: 30 to 300 BPM, accuracy of ±3%
Low battery indicator
Velocity accuracy: ±10% or less comparing with internal phantom testing.
Speaker output:
300 mW or more
External outputs:
Headset, serial port (USB)
Electrical safety:
Conform to IEC60601-1
Internally powered equipment
Type BF applied part.
Operating environment:
10 to 37 °C
85% humidity or less with no condensation