Manual:
Do not soak probes into medicinal solution. Wipe any
contamination from probes with damp cloth.
Disinfection:
Not applicable
Sterilization:
Sterilizable probes except ACP probe:
Low temperature plasma sterilization (Hydrogen
peroxide low temperature plasma sterilization), under
60°C.
Sterilization system is compatible with only the
STERRAD
®
by Johnson & Johnson, K.K. sterilization
system as follows:
・
STERRAD
®
50
・
STERRAD
®
100S (only short cycle)
・
STERRAD
®
200 (only short cycle)
・
STERRAD
®
NX (only standard cycle)
・
STERRAD
®
100NX (only standard cycle)
Do not put liquid, powder & cellulose inside sterilization
equipment or it may reduce effectiveness of sterilization
because these substances absorb hydrogen peroxide.
Eliminate water on surface of probe because it may
reduce effectiveness of sterilization.
Sterilization should be performed in accordance with
instructions of the sterilization equipment.
ACP probe: Steam autoclave
• 30 minutes at 121 °C
• 4 minutes at 134 °C
Do not expose probes to temperatures exceeding 134 °C
Sterilization should be performed in accordance with
instructions of the sterilization equipment.
Drying:
Sterilizable probes except ACP probe:
No particular requirements.
ACP probe: Dry it well after the sterilization.
Maintenance: No particular requirements.
Inspection and Function Testing:
No cracks nor contaminations in appearance.