Logging and Data Transfer

Surveyor 4a File Types (with Memory)

Page 23

Print-Ready will send the file to a printer or create a text file. 

SS-Importable will create a spreadsheet-importable file. 

XMODEM:Binary if for factory diagnostic purposes only. 

If Printer-Ready is select:

a.

Choose the type of statistics which will append to the print-ready output. Select NoStats, 
Daily Stats, Total Stats, or Both Stats and press SELECT. 

NoStats—

Daily Stats—Information on about the files statistics for each day will be sent to the printer 
and/or capture file. Information regarding the number of readings, the minimum value and 
the time it was recorded, the maximum value and the time it was recorded, the maximum 
change between two points and the time it was recorded, the mean, and the standard 
deviation for the day’s data will be displayed. 

Total Stats—Information is similar to Daily Status, except Total Stats provides the overall 
statistics over the entire logging run. 

Both Stats—Daily and overall statistics are calculated. 

b.

Turn the printer on or open a capture file. Then press any key. 

c.

A transfer process screen will appear on the Surveyor 4a screen to show the file transfer is 
taking place. 

d.

After the transfer is complete, a "Deactive printer and/or close capture file" message will 
appear. 

e.

Turn the printer off or close the capture file. 

Note: 

At the bottom of the file, a "Recovery finished line" will be displayed. 

If SS-Importable Transmission is selected:

a.

After selecting SS-Importable, a "To start XMODEM transfer...Press any key..." message 
will be displayed. Press any key to continue. 

b.

Follow the communication software prompts. Press ESC on the computer keyboard or the 
Abort key form the Surveyor 4a to cancel the file transfer. 

c.

When the file transfer is complete, a "Transfer completed!" message will be displayed. 
Press any key to continue. 

d.

To view the transmitted file, go to the computer directory where the file was saved and 
open the file. 

4.3.9.2 Captured and Downloaded Text Transmission

Note: 

Prior to starting the transmit process, please refer to your communications program specific 

commands. 

1.

Select Files from the Main Menu. 

2.

Highlight Transmit and press SELECT.

3.

Highlight the captured text file and press SELECT.

4.

Select ASCII: Text, XMODEM: Text, or XMODEM: Binary. 

Содержание Surveyor 4a

Страница 1: ...003070HY Surveyor 4a USER MANUAL Hach Company 2004 All rights reserved Printed in the U S A eac dp 12 04 Edition 2...

Страница 2: ......

Страница 3: ...Value 13 Section 3 Sonde Setup 15 3 1 Selecting Parameters 15 3 2 Turning the Circulator On Off 15 3 4 Assigning the SDI 12 Address 15 3 5 Setting the SDI Delay 16 3 6 Assigning a Site ID 16 3 7 Assi...

Страница 4: ...ry 25 5 3 Battery Replacement 26 5 3 1 Internal Battery Replacement 26 5 3 2 Lithium Battery Replacement 27 5 4 Calibration 28 Section 6 Troubleshooting 29 6 1 Power Cables and Connections 29 6 2 Inte...

Страница 5: ...y options Item Number Name Description 1 Screen The high resolution liquid crystal screen displays real time reading graphs or submenus 2 Function Keys Takes the action indicated on one of the four re...

Страница 6: ...to the interface mode selected from the internal setup menu The history line also displays the path selected when moving through the menu and submenu structure 1 Real time Parameters 3 Function Key Op...

Страница 7: ...word Allows Access To Level 0 Not Required Surveyor 4a security level at power up Power On Off Backlight On Off Surveyor4 Installed Options Display Real time Tabular Data Screen Real time Graph Data S...

Страница 8: ...Visit us at www hachenvironmental com...

Страница 9: ...e connections to the Surveyor 4a 6 Power the Surveyor 4a on The history line will reflect the interface mode For all modes except Terminal the Surveyor 4a will try each baud rate available starting wi...

Страница 10: ...Setting the Tabular Display Mode The Tabular display mode is the default mode when the instrument is powered on The data is displayed in lines or columns depending on the size of the characters and th...

Страница 11: ...yor 4a shuts down power it back up using the OFF ON key 1 From the Main Screen select SETUP CAL SETUP 2 Highlight Timeout Shutdown using the cursor keys and press SELECT 3 A Timeout Shutdown screen is...

Страница 12: ...ect SETUP CAL SETUP 1 Highlight Timeout Backlight using the cursor keys and press SELECT 2 The Timeout Backlight screen will be displayed 3 Choose the shutdown interval in seconds 1 255 using the curs...

Страница 13: ...Factory Options using the cursor keys and press SELECT 3 Enter the Level 3 password and press DONE 4 Highlight Barometer GPS and press Select 5 Enter which options to install 0 None 1 BP only 2 GPS on...

Страница 14: ...Visit us at www hachenvironmental com...

Страница 15: ...e parameters included on the Sonde 3 2 Turning the Circulator On Off Note When using a DataSonde 3 or Recorder this option allows the stirrer to be turned on of off 1 Highlight Circltr ON Off and pres...

Страница 16: ...y 3 Press DONE when complete 3 7 Assigning the MODBUS Address 1 Highlight MODBUS Address and press SELECT 2 Enter the new Modbus address 1 247 using the cursor keys and then the SELECT key 3 Press DON...

Страница 17: ...Press the DONE key and then the GOBACK key three times 4 Press the Store key One scan has been saved to the Surveyor 4a memory 5 If a Clipboard is full messages is displayed refer to 4 2 2 Reviewing...

Страница 18: ...does not have the capacity to receive files Call Technical Support to inquire about upgrades to the Surveyor 4a Note If the Surveyor 4a is connected to a multiprobe with logging you can also go yo So...

Страница 19: ...ipe delete or autolog and press SELECT Note In PC mode Transmit to transfer files from the Surveyor 4a to a PC is added to the menu In terminal mode Download to transfer files from the Series 3 multip...

Страница 20: ...ecord accurate data Warm up time will vary according to the sensors being used and the field conditions The value must be 30 seconds or more When complete press the DONE key 5 A screen of parameters a...

Страница 21: ...2 Choose the file to review and press the Select key 3 Select how to review the contents of the file either from top Beginning or from a specific date and time a If Beginning is selected the informati...

Страница 22: ...ed to the Surveyor 4a before using the AutoLog feature 4 3 9 Transmitting a File for the Surveyor 4a to a Computer This transmission configuration involves connecting the Surveyor 4a to the computer w...

Страница 23: ...or open a capture file Then press any key c A transfer process screen will appear on the Surveyor 4a screen to show the file transfer is taking place d After the transfer is complete a Deactive print...

Страница 24: ...s a parameter was selected both C and F will be included in the file If SpCond mS cm was selected SpCond S cm Res and TDS are included the file If Depth was selected both Depth Meters and Depth Feet w...

Страница 25: ...r to the other end of the charger cable For the 220 power adapter connect the country s corresponding power cord to the IEC 320 connector 2 Turn the Surveyor 4a on 3 A low external power supply screen...

Страница 26: ...screws completely out 4 Lift and remove the back panel from the bottom of the Surveyor 4a 5 Remove the old batteries from the compartment 6 Insert the new battery aligning the notches on the battery...

Страница 27: ...so the front panel faces the work surface 2 Unscrew the four retaining screws securing the top cover of the back panel with a Phillips screwdriver 3 Slowly pull the cover and flip it back on the botto...

Страница 28: ...ometer The Surveyor 4a can calibrate a multiprobe just like with a computer Proper maintenance of all sensors should proceed calibration Refer to the Sonde user guide for detailed sensor maintenance p...

Страница 29: ...lled Check the battery polarity and voltage 6 2 Internal Components To avoid damage to the internal components when opening the Surveyor 4a make sure that the instrument is clean and dray and that all...

Страница 30: ...w the Surveyor 4a is connected NoConn Series4Sonde Series3Sonde Charger PC or Terminal Depending on the operation to perform choose one of the connection types Check cables and connections Check baud...

Страница 31: ...h parameters have been removed from the tabular display Add more parameters to the tabular display Invalid or incomplete text Date Time or Files Too few characters entered or invalid entry Use the BAC...

Страница 32: ...Visit us at www hachenvironmental com...

Отзывы: