background image

Setting the Backlight Feature
Page 12

Surveyor 4a Setup

4.

Enter when the screen save mode will become active. Select 0 (zero) to disable the screen 
saver mode or select the appropriate time (10 to 255 seconds) for the specific operation using 
the 

RIGHT

 and 

LEFT

 cursor keys. Press 

SELECT

 after each number. 

Note: 

If a mistake is made, clear entries one by one using the 

BACKSPACE

 key. 

5.

When all the numbers are entered, press the 

DONE

 key. A “Setup:Successful! Press any key” 

message will be displayed. 

2.8 Setting the Backlight Feature

The backlight feature turns off the backlight after a certain duration of time. This feature conserves 
power in case the Surveyor 4a is not manually powered off. If the backlight auto-shutoff feature is 
activated while using the Surveyor 4a, up to 40% more battery life can be saved than if the option 
was disabled. 

1.

From the Main Screen, select 

SETUP/CAL

>

SETUP

.

1.

Highlight Timeout:Backlight using the cursor keys and press 

SELECT

2.

The Timeout:Backlight screen will be displayed. 

3.

Choose the shutdown interval, in seconds (1–255) using the cursor keys, and then press 
Select. Select 0 (zero) to disable this feature. 

4.

Press Done. A “Setup Successful!” message will appear, press any key to continue. 

2.9 Setting the Baud Rate for the PC

1.

From the Main Screen, select 

SETUP/CAL

>

SETUP

.

2.

Highlight Baudrate:PC using the cursor keys and press 

SELECT

3.

The Baudrate:PC screen will be displayed. 

4.

Use the cursor keys and the 

SELECT

 key to enter the baud rate that the computer uses to 

communicate with the instrument.The baud rate is dependent on the communications program 
that is used. 

5.

Press the 

DONE

 key when complete. 

2.10 Setting the Baud Rate for the Surveyor 4a

1.

From the Main Screen, select 

SETUP/CAL

>

SETUP

.

2.

Highlight Baudrate:Terminal using the cursor keys and press 

SELECT

3.

Use the cursor keys and the 

SELECT

 key to set the Surveyor 4a baud rate (300, 1200, 2400, 

4800, or 9600).

4.

Press the 

DONE

 key when complete. 

Содержание Surveyor 4a

Страница 1: ...003070HY Surveyor 4a USER MANUAL Hach Company 2004 All rights reserved Printed in the U S A eac dp 12 04 Edition 2...

Страница 2: ......

Страница 3: ...Value 13 Section 3 Sonde Setup 15 3 1 Selecting Parameters 15 3 2 Turning the Circulator On Off 15 3 4 Assigning the SDI 12 Address 15 3 5 Setting the SDI Delay 16 3 6 Assigning a Site ID 16 3 7 Assi...

Страница 4: ...ry 25 5 3 Battery Replacement 26 5 3 1 Internal Battery Replacement 26 5 3 2 Lithium Battery Replacement 27 5 4 Calibration 28 Section 6 Troubleshooting 29 6 1 Power Cables and Connections 29 6 2 Inte...

Страница 5: ...y options Item Number Name Description 1 Screen The high resolution liquid crystal screen displays real time reading graphs or submenus 2 Function Keys Takes the action indicated on one of the four re...

Страница 6: ...to the interface mode selected from the internal setup menu The history line also displays the path selected when moving through the menu and submenu structure 1 Real time Parameters 3 Function Key Op...

Страница 7: ...word Allows Access To Level 0 Not Required Surveyor 4a security level at power up Power On Off Backlight On Off Surveyor4 Installed Options Display Real time Tabular Data Screen Real time Graph Data S...

Страница 8: ...Visit us at www hachenvironmental com...

Страница 9: ...e connections to the Surveyor 4a 6 Power the Surveyor 4a on The history line will reflect the interface mode For all modes except Terminal the Surveyor 4a will try each baud rate available starting wi...

Страница 10: ...Setting the Tabular Display Mode The Tabular display mode is the default mode when the instrument is powered on The data is displayed in lines or columns depending on the size of the characters and th...

Страница 11: ...yor 4a shuts down power it back up using the OFF ON key 1 From the Main Screen select SETUP CAL SETUP 2 Highlight Timeout Shutdown using the cursor keys and press SELECT 3 A Timeout Shutdown screen is...

Страница 12: ...ect SETUP CAL SETUP 1 Highlight Timeout Backlight using the cursor keys and press SELECT 2 The Timeout Backlight screen will be displayed 3 Choose the shutdown interval in seconds 1 255 using the curs...

Страница 13: ...Factory Options using the cursor keys and press SELECT 3 Enter the Level 3 password and press DONE 4 Highlight Barometer GPS and press Select 5 Enter which options to install 0 None 1 BP only 2 GPS on...

Страница 14: ...Visit us at www hachenvironmental com...

Страница 15: ...e parameters included on the Sonde 3 2 Turning the Circulator On Off Note When using a DataSonde 3 or Recorder this option allows the stirrer to be turned on of off 1 Highlight Circltr ON Off and pres...

Страница 16: ...y 3 Press DONE when complete 3 7 Assigning the MODBUS Address 1 Highlight MODBUS Address and press SELECT 2 Enter the new Modbus address 1 247 using the cursor keys and then the SELECT key 3 Press DON...

Страница 17: ...Press the DONE key and then the GOBACK key three times 4 Press the Store key One scan has been saved to the Surveyor 4a memory 5 If a Clipboard is full messages is displayed refer to 4 2 2 Reviewing...

Страница 18: ...does not have the capacity to receive files Call Technical Support to inquire about upgrades to the Surveyor 4a Note If the Surveyor 4a is connected to a multiprobe with logging you can also go yo So...

Страница 19: ...ipe delete or autolog and press SELECT Note In PC mode Transmit to transfer files from the Surveyor 4a to a PC is added to the menu In terminal mode Download to transfer files from the Series 3 multip...

Страница 20: ...ecord accurate data Warm up time will vary according to the sensors being used and the field conditions The value must be 30 seconds or more When complete press the DONE key 5 A screen of parameters a...

Страница 21: ...2 Choose the file to review and press the Select key 3 Select how to review the contents of the file either from top Beginning or from a specific date and time a If Beginning is selected the informati...

Страница 22: ...ed to the Surveyor 4a before using the AutoLog feature 4 3 9 Transmitting a File for the Surveyor 4a to a Computer This transmission configuration involves connecting the Surveyor 4a to the computer w...

Страница 23: ...or open a capture file Then press any key c A transfer process screen will appear on the Surveyor 4a screen to show the file transfer is taking place d After the transfer is complete a Deactive print...

Страница 24: ...s a parameter was selected both C and F will be included in the file If SpCond mS cm was selected SpCond S cm Res and TDS are included the file If Depth was selected both Depth Meters and Depth Feet w...

Страница 25: ...r to the other end of the charger cable For the 220 power adapter connect the country s corresponding power cord to the IEC 320 connector 2 Turn the Surveyor 4a on 3 A low external power supply screen...

Страница 26: ...screws completely out 4 Lift and remove the back panel from the bottom of the Surveyor 4a 5 Remove the old batteries from the compartment 6 Insert the new battery aligning the notches on the battery...

Страница 27: ...so the front panel faces the work surface 2 Unscrew the four retaining screws securing the top cover of the back panel with a Phillips screwdriver 3 Slowly pull the cover and flip it back on the botto...

Страница 28: ...ometer The Surveyor 4a can calibrate a multiprobe just like with a computer Proper maintenance of all sensors should proceed calibration Refer to the Sonde user guide for detailed sensor maintenance p...

Страница 29: ...lled Check the battery polarity and voltage 6 2 Internal Components To avoid damage to the internal components when opening the Surveyor 4a make sure that the instrument is clean and dray and that all...

Страница 30: ...w the Surveyor 4a is connected NoConn Series4Sonde Series3Sonde Charger PC or Terminal Depending on the operation to perform choose one of the connection types Check cables and connections Check baud...

Страница 31: ...h parameters have been removed from the tabular display Add more parameters to the tabular display Invalid or incomplete text Date Time or Files Too few characters entered or invalid entry Use the BAC...

Страница 32: ...Visit us at www hachenvironmental com...

Отзывы: