Surveyor 4a File Types (with Memory)
Page 18

Logging and Data Transfer

XMODEM: Binary is for factory diagnostic purposes and should not be used by the 
operator. This option requires Security Level 3. 

Note: 

Refer to 

section 2.13 on page 9

 to correctly configure the spreadsheet-importable format. 

Before starting the transmit process, refer to the communications program specific commands or 
to ProComm Plus for DOS basic commands or Hyperterminal basic command in the Quick 
Reference chapter. 

5.

Launch the communication program on the PC. Make sure the correct baud rate is selected for 
the computer and that the proper transfer protocol is selected in the communications program 
setup menus. 

6.

From the PC, select the option to receive the files in the communications program. For 
example, press PgDn for ProComm Plus for DOS. Name the file and press ENTER. Then 
press any key. 

7.

The last two lines (with zeros) on the Surveyor 4a screen should start changing to show that 
the transfer is taking place. Otherwise, check the cables, setup, and transfer procedures. 
Press ABort to terminate the transfer.

8.

When the transfer is complete, go to the computer directory where the files was saved. Open 
the files to view the scan(s). 

Note: 

Information from the Series 3 or 4 mulitprobes cannot be downloaded since the Surveyor 4a Clipboard 

memory does not have the capacity to receive files. Call Technical Support to inquire about upgrades to 
the Surveyor 4a. 

Note: 

If the Surveyor 4a is connected to a multiprobe with logging, you can also go yo Sonde instead of 

Surveyor 4a from the Files submenu to create, delete, or look at the status of the multiprobe files. To 
download and file, a Surveyor 4a with memory must be sued. Refer to XXXXX

4.2.4 Deleting Scans

To delete scans stored in the instrument clipboard, follow the steps below.

1.

Press the Files key, or the Files key and then Surveryor4a key (for the Series4Sonde or 
Series4Sonde connections), move the cursor to Delete and press SELECT. 

2.

Make sure you want to delete the selected scans. There is not an undo function. Either leave 
the cursor on All: Date—Time to delete all the scans, or move the cursor down to select a 
specific scan and press SELECT. 

4.3 Surveyor 4a File Types (with Memory)

The Surveyor 4a can be set up to store, create, delete, capture, review, annotate, download, 
transmit, wipe, and view the status of files, and turn the Autolog feature on or off. 

Surveyor File Types

Captured Text: this files is a text file that was created, then data was added manually by 
pressing the STORE key in NoConn, Series3Sonde, or Series4Sonde mode. Once transferred 
to a PC, the captured text can be opened in a word processing program.

Manual File: this files was created using a Series 3 multiprobe in Series3Sonde interface 
mode. Once transferred to a PC, a manual file can be opened in a word processing or 
spreadsheet program. 

Содержание Surveyor 4a

Страница 1: ...003070HY Surveyor 4a USER MANUAL Hach Company 2004 All rights reserved Printed in the U S A eac dp 12 04 Edition 2...

Страница 2: ......

Страница 3: ...Value 13 Section 3 Sonde Setup 15 3 1 Selecting Parameters 15 3 2 Turning the Circulator On Off 15 3 4 Assigning the SDI 12 Address 15 3 5 Setting the SDI Delay 16 3 6 Assigning a Site ID 16 3 7 Assi...

Страница 4: ...ry 25 5 3 Battery Replacement 26 5 3 1 Internal Battery Replacement 26 5 3 2 Lithium Battery Replacement 27 5 4 Calibration 28 Section 6 Troubleshooting 29 6 1 Power Cables and Connections 29 6 2 Inte...

Страница 5: ...y options Item Number Name Description 1 Screen The high resolution liquid crystal screen displays real time reading graphs or submenus 2 Function Keys Takes the action indicated on one of the four re...

Страница 6: ...to the interface mode selected from the internal setup menu The history line also displays the path selected when moving through the menu and submenu structure 1 Real time Parameters 3 Function Key Op...

Страница 7: ...word Allows Access To Level 0 Not Required Surveyor 4a security level at power up Power On Off Backlight On Off Surveyor4 Installed Options Display Real time Tabular Data Screen Real time Graph Data S...

Страница 8: ...Visit us at www hachenvironmental com...

Страница 9: ...e connections to the Surveyor 4a 6 Power the Surveyor 4a on The history line will reflect the interface mode For all modes except Terminal the Surveyor 4a will try each baud rate available starting wi...

Страница 10: ...Setting the Tabular Display Mode The Tabular display mode is the default mode when the instrument is powered on The data is displayed in lines or columns depending on the size of the characters and th...

Страница 11: ...yor 4a shuts down power it back up using the OFF ON key 1 From the Main Screen select SETUP CAL SETUP 2 Highlight Timeout Shutdown using the cursor keys and press SELECT 3 A Timeout Shutdown screen is...

Страница 12: ...ect SETUP CAL SETUP 1 Highlight Timeout Backlight using the cursor keys and press SELECT 2 The Timeout Backlight screen will be displayed 3 Choose the shutdown interval in seconds 1 255 using the curs...

Страница 13: ...Factory Options using the cursor keys and press SELECT 3 Enter the Level 3 password and press DONE 4 Highlight Barometer GPS and press Select 5 Enter which options to install 0 None 1 BP only 2 GPS on...

Страница 14: ...Visit us at www hachenvironmental com...

Страница 15: ...e parameters included on the Sonde 3 2 Turning the Circulator On Off Note When using a DataSonde 3 or Recorder this option allows the stirrer to be turned on of off 1 Highlight Circltr ON Off and pres...

Страница 16: ...y 3 Press DONE when complete 3 7 Assigning the MODBUS Address 1 Highlight MODBUS Address and press SELECT 2 Enter the new Modbus address 1 247 using the cursor keys and then the SELECT key 3 Press DON...

Страница 17: ...Press the DONE key and then the GOBACK key three times 4 Press the Store key One scan has been saved to the Surveyor 4a memory 5 If a Clipboard is full messages is displayed refer to 4 2 2 Reviewing...

Страница 18: ...does not have the capacity to receive files Call Technical Support to inquire about upgrades to the Surveyor 4a Note If the Surveyor 4a is connected to a multiprobe with logging you can also go yo So...

Страница 19: ...ipe delete or autolog and press SELECT Note In PC mode Transmit to transfer files from the Surveyor 4a to a PC is added to the menu In terminal mode Download to transfer files from the Series 3 multip...

Страница 20: ...ecord accurate data Warm up time will vary according to the sensors being used and the field conditions The value must be 30 seconds or more When complete press the DONE key 5 A screen of parameters a...

Страница 21: ...2 Choose the file to review and press the Select key 3 Select how to review the contents of the file either from top Beginning or from a specific date and time a If Beginning is selected the informati...

Страница 22: ...ed to the Surveyor 4a before using the AutoLog feature 4 3 9 Transmitting a File for the Surveyor 4a to a Computer This transmission configuration involves connecting the Surveyor 4a to the computer w...

Страница 23: ...or open a capture file Then press any key c A transfer process screen will appear on the Surveyor 4a screen to show the file transfer is taking place d After the transfer is complete a Deactive print...

Страница 24: ...s a parameter was selected both C and F will be included in the file If SpCond mS cm was selected SpCond S cm Res and TDS are included the file If Depth was selected both Depth Meters and Depth Feet w...

Страница 25: ...r to the other end of the charger cable For the 220 power adapter connect the country s corresponding power cord to the IEC 320 connector 2 Turn the Surveyor 4a on 3 A low external power supply screen...

Страница 26: ...screws completely out 4 Lift and remove the back panel from the bottom of the Surveyor 4a 5 Remove the old batteries from the compartment 6 Insert the new battery aligning the notches on the battery...

Страница 27: ...so the front panel faces the work surface 2 Unscrew the four retaining screws securing the top cover of the back panel with a Phillips screwdriver 3 Slowly pull the cover and flip it back on the botto...

Страница 28: ...ometer The Surveyor 4a can calibrate a multiprobe just like with a computer Proper maintenance of all sensors should proceed calibration Refer to the Sonde user guide for detailed sensor maintenance p...

Страница 29: ...lled Check the battery polarity and voltage 6 2 Internal Components To avoid damage to the internal components when opening the Surveyor 4a make sure that the instrument is clean and dray and that all...

Страница 30: ...w the Surveyor 4a is connected NoConn Series4Sonde Series3Sonde Charger PC or Terminal Depending on the operation to perform choose one of the connection types Check cables and connections Check baud...

Страница 31: ...h parameters have been removed from the tabular display Add more parameters to the tabular display Invalid or incomplete text Date Time or Files Too few characters entered or invalid entry Use the BAC...

Страница 32: ...Visit us at www hachenvironmental com...

Отзывы: