Conductivity Menu Options
Table 2 How to Change Conductivity Menu Options
Setup
How to Get There
Setup 5
Display Lock
From conductivity reading mode, press
SETUP
.
Press the
UP ARROW
until the Setup number is
5
.
Press
ENTER
to toggle the display lock off and on.
When the desired option is selected, press
EXIT
to return to the reading mode.
Setup 6
Temperature
Correction Value
From conductivity reading mode, press
SETUP
.
Press the
UP ARROW
until the Setup number is
6
.
Press
ENTER
.
To select the non-linear function, press the
UP ARROW
or
DOWN ARROW
until the
display shows
NaCl
. Press
ENTER
to accept the setting.
To select a linear conversion, scroll until the correction coefficient appears
(e.g., 2.000%).
Enter the desired value using the numeric keypad. Press
ENTER
to accept the
value. If a number entry error occurs, start over by pressing
CE
.
Note:
If the compensation factor is set to 0.00%, the conductivity readings will not be
corrected for temperature.
When the desired option is selected, press
EXIT
to return to the reading mode.
Setup 7
TDS Correction
Factor
From conductivity reading mode, press
SETUP
.
Press the
UP ARROW
until the Setup number is
7
.
Press
ENTER
.
Press the
UP
or
DOWN ARROW
keys to switch between linear and non-linear
correction functions.
To choose a non-linear conversion, scroll until a flashing
NaCl
appears, then
press
ENTER
.
To select a linear conversion, scroll until the conversion coefficient appears. Use
the numeric keypad to set the value of the coefficient, then press
ENTER
. If an
number entry error occurs, start over by pressing
CE
.
When the desired option is selected, press
EXIT
to return to the reading mode.
Setup 8
Auto Print
Intervals
From conductivity reading mode, press
SETUP
.
Press the
UP ARROW
until the Setup number is
8
.
Press
ENTER
.
Change the print interval by pressing the
UP ARROW
or
DOWN ARROW
.
Press
ENTER
to accept the print interval.
Setup 9
Cond. Standard
Reference
Temperature
From conductivity reading mode, press
SETUP
.
Press the
UP ARROW
until the Setup number is
9
.
Press
ENTER
to toggle between 20° C and 25° C. The default is 25° C.
Press
EXIT
to return to the reading mode.
Setup 10
Temperature
Correction
Off and On
From conductivity reading mode, press
SETUP
.
Press the
UP ARROW
until the Setup number is
10
.
Press
ENTER
to toggle between off and on.
Press
EXIT
to return to the reading mode.