Dissolved Oxygen Calibration in Water Saturated Air

1.

Plug in the DO probe at least 30 minutes before calibrating to
polarize the probe.

2.

Secure the calibration and storage chamber to the probe cable.

3.

Prepare the calibration and storage chamber by holding it under
water and squeezing it a couple of times to pull a small amount of
water into the lower chamber through the inlet. Alternately, open
the bottom of the chamber and insert a water-soaked sponge.

Note:

Avoid completely filling the lower chamber with water.

4.

Insert the DO probe into the calibration and storage chamber. The
tip of the probe must not be flooded with water or be holding a drop
of water on the membrane.

5.

Allow at least ten minutes for the atmosphere in the chamber to
reach a steady state.

Note:

Gently squeezing the lower chamber a couple of times to force water

saturated air into the probe chamber will speed up stabilization.

Note:

Keep the DO probe at a uniform temperature. When holding the probe,

do not touch the metallic button on the side of the probe. The button is a
thermistor that senses temperature. An inaccurate calibration will result if
the temperature of the thermistor is different from the probe membrane.

6.

Press the

DO

key to put the meter in DO reading mode.

7.

Press the

CAL

key located in the lower left corner of the keypad.

8.

The display will show

100%

. Press the

ENTER

key. The stabilizing

icon will appear while the meter completes the calibration.

9.

When the calibration is complete, the meter will return to reading
mode. Press the

EXIT

key during the calibration sequence to back

out of the calibration routine, one screen at a time, without
completing a calibration.

Содержание SensION156

Страница 1: ...54650 23 and Multiparameter Meters ...

Страница 2: ...esired option is selected press EXIT to return to the reading mode Setup 5 Display Lock From pH reading mode press SETUP Press the UP ARROW until the Setup number is 5 Press ENTER to toggle display lock off and on When the desired option is selected press exit to return to the reading mode Setup 6 Measurement Resolution From pH reading mode press SETUP Press the UP ARROW until the Setup number is ...

Страница 3: ... by three beeps If measuring in mV mode the three beeps will still sound when the stabilization occurs but ENTER must be pressed to accept the reading This lets the operator control the acceptance point of the buffer 5 When the reading has stabilized or been accepted the standard number will change to 2 6 Remove the probe from the first buffer and rinse with deionized water Place the probe into th...

Страница 4: ...onductivity reading mode press SETUP Press the UP ARROW until the Setup number is 7 Press ENTER Press the UP or DOWN ARROW keys to switch between linear and non linear correction functions To choose a non linear conversion scroll until a flashing NaCl appears then press ENTER To select a linear conversion scroll until the conversion coefficient appears Use the numeric keypad to set the value of th...

Страница 5: ... the most recent type of calibration 5 Look at the units field to see what kind of calibration is active Scroll to the preferred units using the UP or DOWN ARROWS 6 Use the number keys to change the numeric value if desired It is not necessary to fill up the numeric entry screen before moving on To clear the numeric display press CE 7 When the value and units are correct press ENTER to calibrate o...

Страница 6: ...press EXIT to return to the reading mode Setup 7 Salinity Factor Determine sample salinity The units for salinity are parts per thousand 0 00 From dissolved oxygen reading mode press SETUP Press the UP ARROW until the Setup number is 7 Press ENTER Use the number keys to enter a salinity factor ranging from 0 to 42 Press ENTER to accept the value or EXIT to leave the value unchanged When the desire...

Страница 7: ...n the chamber to reach a steady state Note Gently squeezing the lower chamber a couple of times to force water saturated air into the probe chamber will speed up stabilization Note Keep the DO probe at a uniform temperature When holding the probe do not touch the metallic button on the side of the probe The button is a thermistor that senses temperature An inaccurate calibration will result if the...

Страница 8: ...rily disconnect the probe If the CAL and icons flash after calibrating the DO probe the probe is not polarized Connect the pH electrode to the black connector and the DO or conductivity electrodes to the blue connector If the meter is not in reading mode it may take a few key presses to turn the meter off If the readings do not change when you move an electrode to a new sample the meter probably h...

Отзывы: