5
Conductivity Calibration with a Known Standard
1.
Make sure the meter is in Conductivity reading mode.
2.
Make sure that the reference temperature in Conductivity Setup 9
matches the reference temperature of the sample.
3.
Place the probe in a conductivity standard that is in the expected
range of the samples. Agitate the probe to dislodge bubbles in the
cell. Avoid resting the probe on the bottom or side of the container.
4.
Press
CAL
. The meter will recall the most recent type of calibration.
5.
Look at the units field to see what kind of calibration is active.
Scroll to the preferred units using the
UP
or
DOWN ARROWS
.
6.
Use the number keys to change the numeric value, if desired. It is
not necessary to fill up the numeric entry screen before moving on.
To clear the numeric display, press
CE
.
7.
When the value and units are correct, press
ENTER
to calibrate on
the standard. The meter automatically corrects the calibration
measurement to the selected reference temperature using the NaCl-
based, non-linear temperature coefficient.
8.
The meter will return to Conductivity reading mode when the
calibration is finished.